MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/17571774261477869

Медицинская библиотека · Journal of infection prevention · 2026

Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.

Gupta Sunit Kumar, Aggarwal Nidhima, P Nisha M, Chamala Vamshidhar, Dey Samarjit, V Eeshwar M, Bansal Sumit, Tripathi Mukesh, Chawla Jitendra

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Healthcare workers in critical care settings are at increased risk of acquiring respiratory infections, particularly during aerosol-generating procedures (AGPs). Portable negative-pressure isolation devices have been proposed to fix airborne infection in isolation rooms.

OBJECTIVE

To evaluate whether an isolated negative-pressure device reduces bacterial and fungal aerosol dispersion under controlled laboratory conditions.

METHODS

This laboratory-based controlled experimental observational study assessed aerosol containment across three techniques: (1) no cover, (2) plastic sheet canopy barrier, and (3) a portable negative-pressure isolation system. Standardized 5 mL suspensions of Staphylococcus epidermidis and Aspergillus flavus were nebulized on three separate days for each technique. Nutrient agar plates were placed and colony-forming units (CFU) were manually counted. The primary outcome was the mean CFU per technique per day. One-way ANOVA with post-hoc Welch's t-tests (Bonferroni-corrected) was used for statistical comparison.

RESULTS

A total of 36 bacterial and fungal plates were analysed. For bacterial aerosols, mean CFU was highest with no cover (535.6 ± 5.38), lower with plastic sheet canopy barrier (429.5 ± 5.41), and lowest with the negative-pressure system (217.5 ± 17.5). Differences were statistically significant (ANOVA p = 9.8 × 10-8; η2 = 0.995). Fungal CFU demonstrated a similar pattern with significantly lower CFU for Technique 3 compared with both other techniques (p < .001).

CONCLUSION

The negative-pressure isolation device significantly reduced airborne dispersion of bacterial and fungal aerosols compared with plastic sheet canopy barrier methods. These findings support the use of portable negative-pressure systems to enhance aerosol containment during AGPs. Further clinical validation is recommended.

Для цитирования: Gupta Sunit Kumar, Aggarwal Nidhima, P Nisha M, Chamala Vamshidhar, Dey Samarjit, V Eeshwar M, Bansal Sumit, Tripathi Mukesh, Chawla Jitendra. Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.. Journal of infection prevention. 2026. DOI: 10.1177/17571774261477869.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.