m+Medelevator.RU
← Поиск публикаций
JACC. Cardiovascular interventions · 2019

1-Year Outcomes of the CENTERA-EU Trial Assessing a Novel Self-Expanding Transcatheter Heart Valve.

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Tchétché Didier, Windecker Stephan, Kasel A Markus, Schaefer Ulrich, Worthley Stephen, Linke Axel, Abdel-Wahab Mohamed, Le Breton Herve, Søndergaard Lars, Spence Mark S, Petronio Sonia, Baumgartner Helmut, Hovorka Tomas, Blanke Philipp, Reichenspurner Hermann

Аннотация

OBJECTIVES: The purpose of this study is to report the 1-year results of the CENTERA-EU trial.

BACKGROUND: The CENTERA transcatheter heart valve (THV) (Edwards Lifesciences, Irvine, California) is a low-profile (14-F eSheath compatible), self-expanding nitinol valve, with a motorized delivery system allowing for repositionability. The 30-day results of the CENTERA-EU trial demonstrated the short-term safety and effectiveness of the valve.

METHODS: Implantations were completed in 23 centers in Europe, Australia, and New Zealand. Transfemoral access was used in all patients. Echocardiographic outcomes were adjudicated by a core laboratory at baseline, discharge, 30 days, 6 months, and 1 year. Major adverse clinical events were adjudicated by an independent clinical events committee.

RESULTS: Between March 2015 and July 2016, 203 high-risk patients (age 82.7 ± 5.5 years, 67.5% women, 68.0% New York Heart Association functional class III or IV, Society of Thoracic Surgeons score 6.1 ± 4.2%) with severe, symptomatic aortic stenosis underwent transcatheter aortic valve replacement with the CENTERA THV. The primary endpoint of the study was 30-day mortality (1.0%). At 1 year, overall mortality was 9.1%, cardiovascular mortality was 4.6%, disabling stroke was 4.1%, new permanent pacemakers were implanted in 6.5% of patients at risk, and cardiac-related rehospitalization was 6.8%. Hemodynamic parameters were stable at 1 year, with a mean aortic valve gradient of 8.1 ± 4.7 mm Hg, a mean effective orifice area of 1.7 ± 0.42 cm2, and no incidences of severe or moderate aortic regurgitation.

CONCLUSIONS: The CENTERA-EU trial demonstrated mid-term safety and effectiveness of the CENTERA THV, with low mortality, sustained improvements in hemodynamic performances, and low incidence of permanent pacemaker implantations in high-risk patients with symptomatic aortic stenosis. (Safety and Performance of the Edwards CENTERA-EU Self-Expanding Transcatheter Heart Valve [CENTERA-2]; NCT02458560).

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты