Validation of the InBody BPBIO210 manual auscultatory hybrid device for professional office use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard.

Ntineri Angeliki, Menti Ariadni, Kyriakoulis Konstantinos G, Bountzona Ioanna, Prapa Sofia, Kollias Anastasios, Stergiou George S

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the accuracy of the InBody BPBIO210 manual auscultatory mercury-free hybrid blood pressure (BP) measuring device for professional office use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018).

METHODS

Subjects were recruited to fulfil the age, sex, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in a general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-32 (medium) and 32-42 cm (large).

RESULTS

A total of 94 subjects were recruited and 86 were analysed (mean age 53.7 ± 18.4 [SD] years, 50 men, arm circumference 32.5 ± 4.8 cm, range 22-42 cm). For the validation Criterion 1, the mean ± SD of the differences between the test device and reference BP readings was -1.0 ± 4.1/-0.7 ± 2.5 mmHg (systolic/diastolic). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 2.65/1.59 mmHg (systolic/diastolic).

CONCLUSIONS

The InBody BPBIO210 manual auscultatory hybrid device for professional office BP measurement fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in a general population and can be recommended for clinical use.

Для цитирования: Ntineri Angeliki, Menti Ariadni, Kyriakoulis Konstantinos G, Bountzona Ioanna, Prapa Sofia, Kollias Anastasios, Stergiou George S. Validation of the InBody BPBIO210 manual auscultatory hybrid device for professional office use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard.. Blood pressure monitoring. 2022. DOI: 10.1097/mbp.0000000000000571.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.