Rapid communications in mass spectrometry : RCM · 2017
Calibration bias of experimentally determined chlorine isotope enrichment factors: the need for a two-point calibration in compound-specific chlorine isotope analysis.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ebert Karin A, Laskov Christine, Elsner Martin, Haderlein Stefan B
Аннотация
RATIONALE: The recent development of compound-specific online chlorine isotope analysis (37 Cl-CSIA) methods has fostered dual chlorine-carbon isotope studies to gain better insights into sources and environmental transformation reactions of chlorinated ethenes. One-point and two-point calibration schemes are currently used to convert raw data to the international δ37 ClSMOC scale, but a critical evaluation of best practices to arrive at reliable δ37 ClSMOC signatures and enrichment factors was missing and is presented here.
METHODS: Aqueous solutions of neat perchloroethylene and trichloroethylene (TCE) and aqueous samples from a TCE biodegradation experiment with pure cultures of Desulfitobacterium hafniense Y51 were analysed for their chlorine isotope ratios using GC/qMS and GC/IRMS. The δ37 ClSMOC values were obtained using one-point and two-point calibration schemes. Chlorine isotope enrichment factors, εCl , were calculated using both approaches and the corresponding bias of δ37 ClSMOC values introduced by the different types of calibration was determined.
RESULTS: Different calibration methods resulted in significant differences (up to 30%) in both δ37 Cl signatures and εCl values.
CONCLUSIONS: Our results demonstrate that a two-point calibration together with comprehensive information on reference materials is indispensable and should become standard practice for reliable 37 Cl-CSIA of organic compounds. Copyright © 2016 John Wiley & Sons, Ltd.
METHODS: Aqueous solutions of neat perchloroethylene and trichloroethylene (TCE) and aqueous samples from a TCE biodegradation experiment with pure cultures of Desulfitobacterium hafniense Y51 were analysed for their chlorine isotope ratios using GC/qMS and GC/IRMS. The δ37 ClSMOC values were obtained using one-point and two-point calibration schemes. Chlorine isotope enrichment factors, εCl , were calculated using both approaches and the corresponding bias of δ37 ClSMOC values introduced by the different types of calibration was determined.
RESULTS: Different calibration methods resulted in significant differences (up to 30%) in both δ37 Cl signatures and εCl values.
CONCLUSIONS: Our results demonstrate that a two-point calibration together with comprehensive information on reference materials is indispensable and should become standard practice for reliable 37 Cl-CSIA of organic compounds. Copyright © 2016 John Wiley & Sons, Ltd.