MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.31002

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2024

Right disc thrombosis of the new Gore Cardioform ASD Occluder.

Oliva Omar A, Giugno Luca, Moroni Alice, Sturla Francesco, Piazza Luciane, Saracino Antonio, Micheletti Angelo, d'Aiello Angelo Fabio, Reali Matteo, Chessa Massimo, Carminati Mario

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The GCA is a well performing device in terms of efficacy despite complex anatomies (aortic rim <5 mm and ASD diameter >17 mm) with a good safety profile.

AIMS: To evaluate atrial septal defect (ASD) features impacting on right disc device thrombosis in patients who underwent Gore Cardioform ASD Occluder (GCA) implantation.

METHODS

A total of 44 consecutive patients undergoing percutaneous ASD with GCA device from January 2020 to September 2022 at our tertiary care Center were evaluated. The minimum follow-up was 6 months.

RESULTS

The patients were stratified in two groups according to a cut-off value of ASD diameter equal to 20 mm at sizing balloon, derived from ROC analysis (AUC = 0.894; p = 0.024). Baseline characteristics were comparable between groups in terms of age, sex, weight, height, and interatrial septum dimensions. Patients with ASD > 20 mm (n = 9) had a higher ASD/device dimension ratio, both at echocardiography (p = 0.009) and at sizing balloon (p = 0.001), longer fluoroscopic time (p = 0.022), and higher incidence of device thrombosis (0.006). Right disc thrombosis was observed in three patients of the ASD > 20 mm group, always in the inferior portion of the right disc. On univariate analysis, ASD diameter at sizing balloon (OR 1.360; p = 0.036) was the only positive predictor of device thrombosis.

CONCLUSIONS

Right disc thrombosis of the GCA device may be under-recognized at follow-up, hence deserving clinical attention, especially in those patients with larger ASD diameters.

Для цитирования: Oliva Omar A, Giugno Luca, Moroni Alice, Sturla Francesco, Piazza Luciane, Saracino Antonio, Micheletti Angelo, d'Aiello Angelo Fabio, Reali Matteo, Chessa Massimo, Carminati Mario. Right disc thrombosis of the new Gore Cardioform ASD Occluder.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2024. DOI: 10.1002/ccd.31002.