MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ju.0000000000004945

Медицинская библиотека · The Journal of urology · 2026

First-In-Human Assessment of the UroActive Electronic Artificial Urinary Sphincter for Treating Male Patients with Stress Urinary Incontinence.

Mozer Pierre, Beaugerie Aurelien, Perrouin-Verbe Marie-Aimée, Denormandie Anne, Cotte Juliette, Plassais Caroline, Tran Stéphanie, Elliott Dan, Chartier-Kastler Emmanuel

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To evaluate the feasibility, safety, and efficacy during the first year of use of the UroActive device in male patients with stress urinary incontinence.

MATERIALS AND

METHODS

This prospective, multicenter, single-arm, first-in-human study enrolled 6 male patients with bothersome stress urinary incontinence unresponsive to conservative or surgical treatments. After UroActive implantation, follow-up was planned for 5 years (postactivation). Study end points included successful device activation, revision, and explantation rates. Other end points assessed 24 hours pad-weight test reduction, pad usage, uroflowmetry, postvoid residual urine, and patient-reported outcomes (International Consultation Incontinence Questionnaires UI Short Form, EQ-5D, I-QOL, Urinary Symptom Profile). Adverse events were collected throughout the study.

RESULTS

All 6 patients underwent successful implantation and device activation (100%). No explants or revisions were required during the 365 days postactivation. At 365 days postactivation, the pad weight was reduced by ≥ 50% in all patients and reduced by ≥ 75% in 5 (83.3%). The median pad use decreased from 3.7 to 1.0 pad per day. Uroflowmetry remained within acceptable ranges, and postvoid residual urine was minimal. Quality of life scores significantly improved across all instruments. Nine device deficiencies were reported, all involving the patient remote control. No unanticipated serious adverse events occurred.

CONCLUSIONS

In men with stress urinary incontinence, the UroActive device demonstrates promising feasibility, safety, and functional outcomes during the 365 days postactivation. Patients experienced substantial symptom improvement and high satisfaction, with no surgical revisions. Larger studies with extended follow-up are needed to confirm these findings and assess long-term device performance.

Для цитирования: Mozer Pierre, Beaugerie Aurelien, Perrouin-Verbe Marie-Aimée, Denormandie Anne, Cotte Juliette, Plassais Caroline, Tran Stéphanie, Elliott Dan, Chartier-Kastler Emmanuel. First-In-Human Assessment of the UroActive Electronic Artificial Urinary Sphincter for Treating Male Patients with Stress Urinary Incontinence.. The Journal of urology. 2026. DOI: 10.1097/ju.0000000000004945.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.