DOI 10.3928/1081-597x-20000101-11
Lower intraoperative flap complication rate with the Hansatome microkeratome compared to the Automated Corneal Shaper.
Аннотация
PURPOSE: The purpose of this study was to retrospectively compare the incidence of intraoperative flap complications, such as partial flaps, donut-shaped flaps, central corneal cuts, and complete caps with the Hansatome and Automated Corneal Shaper (ACS) microkeratomes.
METHODS
All laser in situ keratomileusis (LASIK) procedures performed by a single surgeon with the Hansatome or Automated Corneal Shaper in which intraocular pressure was verified with a pneumotonometer were reviewed.
RESULTS
A total of 90 eyes had LASIK with the ACS microkeratome. Six of the ACS eyes (6.7%) had intraoperative flap complications (4 partial flaps, 1 donut-shaped flap, 0 central corneal cuts, 1 complete cap). Partial flaps and donut-shaped flaps were replaced without laser application and the procedure repeated 2 to 3 months later. Two of these eyes lost 2 lines and one lost 1 line of spectacle-corrected visual acuity at 6 months after repeat LASIK. The eye with the donut-shaped flap was treated with transepithelial photorefractive keratectomy (PRK) and had no change in spectacle-corrected visual acuity at 6 months after PRK. The eye with the complete cap had no change in spectacle-corrected visual acuity after laser ablation. Five hundred ninety-eight (598) eyes had LASIK with the Hansatome microkeratome. Two of the Hansatome eyes (0.3%) had a flap complication (1 partial flap and 1 donut-shaped flap). The first eye retained spectacle-corrected visual acuity at 6 months after repeat LASIK. The second eye had transepithelial PRK to eliminate the donut shaped flap with no loss of spectacle-corrected visual acuity at 6 months after surgery. The difference in flap complications between the two procedures was statistically significant (P < .01). There were no flap displacements following surgery in either group.
CONCLUSION
Intraoperative flap complications are less likely to occur with the Hansatome microkeratome than with the ACS microkeratome.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.