Impact of a modified anti-thrombotic guideline on stroke in children supported with a pediatric ventricular assist device.

Rosenthal David N, Lancaster Chacy A, McElhinney Doff B, Chen Sharon, Stein MaryLyn, Lin Aileen, Doan Lan, Murray Jenna M, Gowan Mary Alice, Maeda Katsuhide, Reinhartz Olaf, Almond Christopher S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Stroke is the most feared complication associated with the Berlin Heart EXCOR pediatric ventricular assist device (VAD), the most commonly used VAD in children, and affects 1 in 3 children. We sought to determine whether a modified anti-thrombotic guideline, involving more intense platelet inhibition and less reliance on platelet function testing, is associated with a lower incidence of stroke.

METHODS

All children supported with the EXCOR at Stanford from 2009 to 2014 were divided into 2 cohorts based on the primary anti-thrombotic guideline used to prevent pump thrombosis: (1) the Edmonton Anti-thrombotic Guideline (EG) cohort, which included children implanted before September 2012 when dual anti-platelet therapy was used with doses titrated to Thromboelastrography/PlateletMapping (TEG/PM); and (2) the Stanford Modified Anti-thrombotic Guideline (SG) cohort, which included children implanted on or after September 2012 when triple anti-platelet therapy was used routinely and where doses were uptitrated to high, weight-based dosing targets, with low-dose steroids administered as needed for inflammation.

RESULTS

At baseline, the EG (N = 16) and SG (N = 11) cohorts were similar. The incidence rate of stroke in the SG cohort was 84% lower than in the EG cohort (0.8 vs 4.9 events per 1,000 days of support, p = 0.031), and 86% lower than in the previous Investigational Device Exemption trial (p = 0.006). The bleeding rate was also lower in the SG cohort (p = 0.015). Target doses of aspirin, clopidogrel and dipyridamole were higher (all p < 0.003), with less dosing variability in the SG cohort than in the EG cohort. There was no difference in adenosine diphosphate inhibition by TEG/PM, but arachidonic acid inhibition was higher in the SG cohort (median 75% vs 39%, p = 0.008).

CONCLUSIONS

Stroke was significantly less common in pediatric patients supported with the Berlin Heart EXCOR VAD using a triple anti-platelet regimen uptitrated to high, weight-based dosing targets as compared with the dual anti-platelet regimen titrated to PM, and without a higher risk of bleeding. Larger studies are needed to confirm these findings.

Для цитирования: Rosenthal David N, Lancaster Chacy A, McElhinney Doff B, Chen Sharon, Stein MaryLyn, Lin Aileen, Doan Lan, Murray Jenna M, Gowan Mary Alice, Maeda Katsuhide, Reinhartz Olaf, Almond Christopher S. Impact of a modified anti-thrombotic guideline on stroke in children supported with a pediatric ventricular assist device.. The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation. 2017. DOI: 10.1016/j.healun.2017.05.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.