Endovascular Proximal Forearm Arteriovenous Fistula for Hemodialysis Access: Results of the Prospective, Multicenter Novel Endovascular Access Trial (NEAT).

Lok Charmaine E, Rajan Dheeraj K, Clement Jason, Kiaii Mercedeh, Sidhu Ravi, Thomson Ken, Buldo George, Dipchand Christine, Moist Louise, Sasal Joanna, NEAT Investigators

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Hemodialysis arteriovenous fistulas (AVFs) are suboptimally used primarily due to problems with maturation, early thrombosis, and patient nonacceptance. An endovascular approach to fistula creation without open surgery offers another hemodialysis vascular access option.

STUDY

DESIGN

Prospective, single-arm, multicenter study (Novel Endovascular Access Trial [NEAT]).

SETTINGS &

PARTICIPANTS

Consecutive adult non-dialysis-dependent and dialysis-dependent patients referred for vascular access creation at 9 centers in Canada, Australia, and New Zealand.

INTERVENTION: Using catheter-based endovascular technology and radiofrequency energy, an anastomosis was created between target vessels, resulting in an endovascular AVF (endoAVF).

OUTCOMES: Safety, efficacy, functional usability, and patency end points.

MEASUREMENTS: Safety as percentage of device-related serious adverse events; efficacy as percentage of endoAVFs physiologically suitable (brachial artery flow ≥ 500mL/min, vein diameter ≥ 4mm) for dialysis within 3 months; functional usability of endoAVFs to provide prescribed dialysis via 2-needle cannulation; primary and cumulative endoAVF patencies per standardized definitions.

RESULTS

80 patients were enrolled (20 roll-in and 60 participants in the full analysis set; the latter are reported). EndoAVFs were created in 98% of participants; 8% had a serious procedure-related adverse event (2% device related). 87% were physiologically suitable for dialysis (eg, mean brachial artery flow, 918mL/min; endoAVF vein diameter, 5.2mm [cephalic vein]). EndoAVF functional usability was 64% in participants who received dialysis. 12-month primary and cumulative patencies were 69% and 84%, respectively.

LIMITATIONS: Due to the unique anatomy and vessels used to create endoAVFs, this was a single-arm study without a surgical comparator.

CONCLUSIONS

An endoAVF can be reliably created using a radiofrequency magnetic catheter-based system, without open surgery and with minimal complications. The endoAVF can be successfully used for hemodialysis and demonstrated high 12-month cumulative patencies. It may be a viable alternative option for achieving AVFs for hemodialysis patients in need of vascular access.

Для цитирования: Lok Charmaine E, Rajan Dheeraj K, Clement Jason, Kiaii Mercedeh, Sidhu Ravi, Thomson Ken, Buldo George, Dipchand Christine, Moist Louise, Sasal Joanna, NEAT Investigators. Endovascular Proximal Forearm Arteriovenous Fistula for Hemodialysis Access: Results of the Prospective, Multicenter Novel Endovascular Access Trial (NEAT).. American journal of kidney diseases : the official journal of the National Kidney Foundation. 2017. DOI: 10.1053/j.ajkd.2017.03.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.