First-pass perfusion imaging using cardiovascular magnetic resonance in patients with various cardiac implantable electronic devices.

Meier Claudia, Bietenbeck Michael, Theofanidou Maria, Vehof Volker, Stalling Philipp, Korthals Dennis, Chamling Bishwas, Estepa Misael, Doeblin Patrick, Kelle Sebastian, Yilmaz Ali

СтатьяПеревод в очереди

Аннотация

AIMS: The number of patients with cardiac implantable electronic devices (CIEDs) is increasing. However, there is limited experience regarding vasodilator-stress cardiovascular magnetic resonance (CMR) and resulting device artifacts on perfusion images. The aim of this study was to determine CMR image quality in patients with different CIED types for CMR-based perfusion stress testing.

METHODS AND

RESULTS

A total of 156 patients with active CIEDs underwent CMR on a 1.5-Tesla scanner. Both conventional steady-state-free-precession (SSFP) and modified spoiled gradient-echo (sGE) protocols under stress and resting conditions were used to evaluate image artifacts in a 16-segment segment model of the heart. The study group comprised 39% conventional pacemaker (PM), 4% cardiac resynchronization therapy-pacemaker (CRT-P), 38% conventional implantable cardioverter-defibrillator (ICD), 6% cardiac resynchronization therapy-ICD (CRT-D), and 13% subcutaneous ICD (S-ICD) patients. PM-carriers showed only minor image artifacts in both perfusion protocols. Artifacts caused by ICDs were predominantly located in the left-ventricular (LV) inferolateral and anterior segments. S-ICDs showed the highest extent of artifacts with an anterolateral accentuation. The artifact extent was significantly reduced when sGE-based perfusion was used compared to SSFP-based sequences. 69% of the patients received a stress-perfusion protocol, and elective coronary angiography confirmed the presence of coronary stenosis in three cases. No major safety-relevant issues occurred.

CONCLUSION

Myocardial perfusion imaging by CMR is safe and feasible with moderate-to-high image quality in patients with all types of CIEDs, including non-conditional devices, ICDs, and S-ICDs. A sGE-based perfusion protocol should be preferred in patients with left-sided ICDs, CRT-Ds, or S-ICDs.

Для цитирования: Meier Claudia, Bietenbeck Michael, Theofanidou Maria, Vehof Volker, Stalling Philipp, Korthals Dennis, Chamling Bishwas, Estepa Misael, Doeblin Patrick, Kelle Sebastian, Yilmaz Ali. First-pass perfusion imaging using cardiovascular magnetic resonance in patients with various cardiac implantable electronic devices.. Clinical research in cardiology : official journal of the German Cardiac Society. 2025. DOI: 10.1007/s00392-025-02636-1.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.