Peripheral Vasculature Embolization with the Onyx Liquid Embolic System in a Swine Model.

Siskin Gary P, Rochon Paul J, Gandhi Ripal, Cocciolone Austin, Winter Andrew, Patel Rahul S, Golzarian Jafar, Arslan Bulent

СтатьяПеревод в очереди

Аннотация

PURPOSE: To characterize the histologic and inflammatory changes that resulted from peripheral embolization using ethylene vinyl alcohol ([EVOH]; Onyx) in an animal model. This study also assessed the radiopacity of the Onyx after a 1-minute mixing time.

MATERIALS AND

METHODS

Embolization using EVOH alone or in combination with coils/plugs was performed on large-vessel, small-vessel, and very-small-vessel (3-5 mm, 1-3 mm, and <1 mm in diameter, respectively) targets in the peripheral vasculature of Yucatan miniature pigs. Multiple formulations of EVOH were used. Angiography at 1, 3, and 12 months was used to assess embolic migration and vessel occlusion. Pathological evaluations were performed to assess the biological response to Onyx-mediated embolization.

RESULTS

Angiography demonstrated that the radiopacity of EVOH was sufficient for control during deployment in 98.2% of the procedures. Microcatheter removal was successful in all cases, and there was minimal migration of EVOH after delivery. Pathological findings indicated no significant tissue response or vessel trauma after embolization, and the occlusive material (consisting of EVOH, thrombus, and fibrotic tissue) evolved over time. The inflammatory response was minimal to mild, and vessel wall injury was observed in some cases. Complete to near-complete occlusion was achieved at every embolization site at the time of follow-up, although histologic evidence of recanalization was observed in 59.3% of the treated vessels.

CONCLUSIONS

The study demonstrated the effectiveness of EVOH as an embolic agent in the peripheral vascular system with adequate radiopacity, minimal adverse effects, and acceptable long-term histologic outcomes in a swine model.

Для цитирования: Siskin Gary P, Rochon Paul J, Gandhi Ripal, Cocciolone Austin, Winter Andrew, Patel Rahul S, Golzarian Jafar, Arslan Bulent. Peripheral Vasculature Embolization with the Onyx Liquid Embolic System in a Swine Model.. Journal of vascular and interventional radiology : JVIR. 2025. DOI: 10.1016/j.jvir.2025.08.032.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.