MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00347-009-2096-9

Медицинская библиотека · Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft · 2010

[Transpalpebral measurement of axial eye length. Use of contact B-scan sonography].

Martius P, Tech S, Stachs O, Guthoff R F

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: In ophthalmology, it is very often necessary to determine the axial eye length. When standard measuring techniques fail, B-scan sonography may be used. However, coupling problems inhibit exact measurement. The aim of this study was to counter these problems.

PATIENTS AND

METHODS

Correction values were created for the ultrasonic devices (1) Master-Vu (Sonomed), (2) CineScan (Quantel Medical), and (3) System-ABD (Innovative Imaging), including the position of the coupling layer of the hand pieces and the lid thickness. For evaluation of the method, the axial eye lengths of 35 eyes were measured transpalpebrally. These data were compared to those from the (4) IOL Master (Carl Zeiss Meditec). The usefulness of the developed method was shown by two patients.

RESULTS

The correction values were (1) 3.0 mm, (2) 1.5 mm, and (3) 1.5 mm. The medial aberrations to (4) were (1) 0.4+/-0.5 mm, (2) 0.9+/-0.7 mm, and (3) 0.2+/-0.7 mm.

CONCLUSIONS

The introduced method enables transpalpebral eye length measurement and is also applicable in cases of missing compliance or in eyes with an obfuscated optical medium.

Для цитирования: Martius P, Tech S, Stachs O, Guthoff R F. [Transpalpebral measurement of axial eye length. Use of contact B-scan sonography].. Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft. 2010. DOI: 10.1007/s00347-009-2096-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.