MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jhsa.2020.11.013

Медицинская библиотека · The Journal of hand surgery · 2021

Minimally Invasive Intramedullary Screw Versus Plate Fixation for Proximal Phalanx Fractures: A Biomechanical Study.

Miles Megan R, Krul Kevin P, Abbasi Pooyan, Thakkar Madhuli Y, Giladi Aviram M, Means Kenneth R

СтатьяПеревод в очереди

Аннотация

PURPOSE: To compare the maximum interfragmentary displacement of short oblique proximal phalanx (P1) fractures fixed with an intramedullary headless compression screw (IMHCS) versus a plate-and-screws construct in a cadaveric model that generates finger motion via the flexor and extensor tendons of the fingers.

METHODS

We created a 30° oblique cut in 24 P1s of the index, middle, ring, and little fingers for 3 matched pairs of cadaveric hands. Twelve fractures were stabilized with an IMHCS using an antegrade, dorsal articular margin technique at the P1 base. The 12 matched-pair P1 fractures were stabilized with a radially placed 2.0-mm plate with 2 bicortical nonlocking screws on each side of the fracture. Hands were mounted to a frame allowing a computer-controlled, motor-driven, linear actuator powered movement of fingers via the flexor and extensor tendons. All fingers underwent 2,000 full-flexion and extension cycles. Maximum interfragmentary displacement was continuously measured using a differential variable reluctance transducer.

RESULTS

The observed mean displacement differences between IMHCS and plate-and-screws fixation was not statistically significant throughout all time points during the 2,000 cycles. A 2 one-sided test procedure for paired samples confirmed statistical equivalence in fracture displacement between fixation methods at the final 2,000-cycle time point.

CONCLUSIONS

The IMHCS provided biomechanical stability equivalent to plate-and-screws for short oblique P1 fractures at the 2,000-cycle mark in this cadaveric model.

CLINICAL RELEVANCE: Short oblique P1 fracture fixation with an IMHCS may provide adequate stability to withstand immediate postoperative active range of motion therapy.

Для цитирования: Miles Megan R, Krul Kevin P, Abbasi Pooyan, Thakkar Madhuli Y, Giladi Aviram M, Means Kenneth R. Minimally Invasive Intramedullary Screw Versus Plate Fixation for Proximal Phalanx Fractures: A Biomechanical Study.. The Journal of hand surgery. 2021. DOI: 10.1016/j.jhsa.2020.11.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.