DOI 10.3390/bioengineering10101136
Treatment of Severe Pulmonary Regurgitation in Enlarged Native Right Ventricular Outflow Tracts: Transcatheter Pulmonary Valve Replacement with Three-Dimensional Printing Guidance.
Аннотация
BACKGROUND
Severe pulmonary regurgitation (PR) often occurs following a transannular patch repair of tetralogy of Fallot, resulting in an enlarged native right ventricular outflow tract (nRVOT) with varying shapes.
METHODS
We conducted a single-center study with eight patients having severe PR and enlarged nRVOT (diameters ≥ 29 mm). Transcatheter pulmonary valve replacement (TPVR) was performed using the self-expanding PT-Valve. Preoperative evaluation included echocardiography, computed tomography, and magnetic resonance imaging. A 3D-printed model of the nRVOT was used for preoperative assessment. Follow-up data were collected in 1-year follow-up.
RESULTS
PT-Valve was successfully implanted in all patients, resulting in immediate improvement of severe PR. Pulmonary artery diastolic pressure increased significantly (p < 0.001). No deaths or coronary compression occurred during the procedure. Over a 1-year follow-up, no stent displacement or fracture occurred. Only two patients had trace paravalvular leaks. Magnetic resonance imaging revealed a reversal of right ventricular remodeling, with a significant reduction in right ventricular end-diastolic volume index (p < 0.001) and improved right ventricular ejection fraction (p < 0.001). All patients achieved primary endpoints.
CONCLUSION
3D printing-guided PT-Valve implantation in enlarged nRVOT for severe PR is safe and effective, expanding TPVR indications and offering potential treatment for a broader patient population.