A new occlusion device: application of the ArtVentive endoluminal occlusion system (EOS)--first in human clinical trial.

Venbrux Anthony C, Rudakov Leon, Plass Andre, Emmert Maximilian Y, Ebner Adrian

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: The purpose of this study was to determine the safety and efficacy of a new endoluminal occlusion device, ArtVentive endoluminal occlusion system (EOS), to occlude the spermatic vein in symptomatic males with varicoceles.

METHODS

The ArtVentive EOS device has been developed for percutaneous, peripheral occlusion of the peripheral arterial and venous vasculature. The system is comprised of an implantable occlusion device and a delivery catheter. At present, there are two device sizes: (a) size 1 for target vessels ranging between 3.5 and 5.5 mm in diameter, and (b) size 2 for target vessels 5.5-8.5 mm in diameter. The treatment group included six adult males, ages 22-34 years. Nine target vessels were occluded. A total of 20 devices were implanted in six subjects.

RESULTS

The acute occlusion rate at the end of the procedure was 100 % occurring in nine of nine vessels. The spermatic veins of all patients remained occluded on venography at 30 days follow-up. Pain scores related to varicoceles decreased in five of six patients.

CONCLUSIONS

Although we recognize this study is limited, initial experience indicates that the ArtVentive EOS is a safe and effective new device for occlusion of vessels (varicoceles). The device has potential applications in other clinical conditions requiring occlusion of veins or arteries.

Для цитирования: Venbrux Anthony C, Rudakov Leon, Plass Andre, Emmert Maximilian Y, Ebner Adrian. A new occlusion device: application of the ArtVentive endoluminal occlusion system (EOS)--first in human clinical trial.. Cardiovascular and interventional radiology. 2013. DOI: 10.1007/s00270-013-0626-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.