Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
Safety and Efficacy of Pivot-Bridge for Severe Functional Tricuspid Regurgitation: The First-in-Human Experiences.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kim Eun Kyoung, Chon Min-Ku, Kim Hyun-Sook, Park Yong-Hyun, Suh Jon, Park Jae-Hyeong, Lee Jae-Hwan, Choi Young Jin, Shin Eun Seok, Yoon Hyuck-Jun, Yang Dong Hyun, Park Junhui, Lee Seung-Whan, Hahn Joo-Yong
Аннотация
BACKGROUND: The management of severe tricuspid regurgitation (TR) remains challenging. Although several promising transcatheter devices have been developed, procedural complexity and the need for special imaging techniques are hurdles to their widespread clinical application.
AIMS: The study aimed to evaluate the safety and efficacy of a novel atraumatic vertical spacer, Pivot-Bridge, for severe TR.
METHODS: Patients with severe TR who were considered surgical candidates were enrolled. All procedures were guided by fluoroscopy and transthoracic echocardiography (TTE). Baseline characteristics, serious adverse events, and functional status were collected. TTE and cardiac computed tomography (CT) were serially performed for evaluation of TR and right ventricular (RV) remodeling.
RESULTS: Fifteen patients (68 ± 11 years old, 27% women) were enrolled. Procedural success was achieved in all patients (the mean procedural time of 47 ± 24 min). During the device implantation (median 5 days), no serious adverse events occurred. TR was reduced by one grade or more in all patients. In TTE, vena contracta width and effective regurgitant orifice area significantly decreased by Pivot-Bridge (1.6 ± 0.8 to 0.8 ± 0.3 mm; p < 0.001 and 1.2 ± 0.5 to 0.6 ± 0.3 cm2; p < 0.001). RV end diastolic volume also significantly reduced at follow-up CT (320.2 ± 93.7 to 294.4 ± 81.1 mm3; p = 0.026). Ten patients underwent surgery with removal of the device, and five patients were stabilized by medical treatment, in whom the Pivot-Bridge was removed percutaneously.
CONCLUSIONS: The Pivot-Bridge was safe and effective in patients with severe TR. TR reduction was observed in all patients without periprocedural complications or adverse cardiovascular events during device placement.
REGISTRATION: ClinicalTrials.gov number: NCT05854095.
AIMS: The study aimed to evaluate the safety and efficacy of a novel atraumatic vertical spacer, Pivot-Bridge, for severe TR.
METHODS: Patients with severe TR who were considered surgical candidates were enrolled. All procedures were guided by fluoroscopy and transthoracic echocardiography (TTE). Baseline characteristics, serious adverse events, and functional status were collected. TTE and cardiac computed tomography (CT) were serially performed for evaluation of TR and right ventricular (RV) remodeling.
RESULTS: Fifteen patients (68 ± 11 years old, 27% women) were enrolled. Procedural success was achieved in all patients (the mean procedural time of 47 ± 24 min). During the device implantation (median 5 days), no serious adverse events occurred. TR was reduced by one grade or more in all patients. In TTE, vena contracta width and effective regurgitant orifice area significantly decreased by Pivot-Bridge (1.6 ± 0.8 to 0.8 ± 0.3 mm; p < 0.001 and 1.2 ± 0.5 to 0.6 ± 0.3 cm2; p < 0.001). RV end diastolic volume also significantly reduced at follow-up CT (320.2 ± 93.7 to 294.4 ± 81.1 mm3; p = 0.026). Ten patients underwent surgery with removal of the device, and five patients were stabilized by medical treatment, in whom the Pivot-Bridge was removed percutaneously.
CONCLUSIONS: The Pivot-Bridge was safe and effective in patients with severe TR. TR reduction was observed in all patients without periprocedural complications or adverse cardiovascular events during device placement.
REGISTRATION: ClinicalTrials.gov number: NCT05854095.
Предметные рубрики MeSH
HumansFemaleTricuspid Valve InsufficiencyTreatment OutcomeSeverity of Illness IndexMaleCardiac CatheterizationAgedVentricular Function, RightTime FactorsTricuspid ValveMiddle AgedProsthesis DesignRecovery of FunctionHeart Valve Prosthesis ImplantationHeart Valve ProsthesisVentricular RemodelingHemodynamicsAged, 80 and over