Biomechanical comparison of a novel tensioned cable construct versus tension band wiring for transverse patella fracture fixation.

Kagan Benjamin D, Sundet Alec D, Gardner-Morse Mack, Chaidarun Pirapon L, Beynnon Bruce D, Schottel Patrick C

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

PURPOSE: Tension band wiring (TBW) is the most widely accepted method for patella fracture fixation. The purpose of our study was to compare the biomechanical efficacy of a novel cable construct to TBW for the fixation of transverse patella fractures. The tensioned cable construct was hypothesized to have less fracture gapping after cyclic flexion-extension loading and greater ultimate load to failure as compared to TBW.

METHODS

Transverse patellar osteotomies (AO/OTA 34C1.1) were performed on nine pairs of fresh-frozen human cadaveric whole legs (mean age 82.2 years, range 71-101). Treatment with TBW or tensioned cable construct was randomized within each specimen pair. Fracture site displacement was measured after 5000 flexion-extension cycles from 0° to 90° at 0.5 Hz. In load to failure testing, the knee was fixed at 45° of flexion and the quadriceps tendon was pulled proximally at 0.5 mm/sec until patella fixation failure. Comparisons were made using paired t-tests with alpha values of 0.05.

RESULTS

Eight paired specimens completed the cyclic loading. The tensioned cable construct had significantly less fracture gapping than TBW (2.9 vs 10.9 mm; p = 0.020). Seven paired limbs underwent load to failure testing, which revealed no significant difference between the tensioned cable construct and TBW (1551.6 N vs 1664.0 N; p = 0.26).

CONCLUSION

In this study of transverse patella fracture fixation, a tensioned cable construct demonstrated significantly less fracture gapping compared to TBW in response to cyclic loading with no significant difference in load at failure.

Для цитирования: Kagan Benjamin D, Sundet Alec D, Gardner-Morse Mack, Chaidarun Pirapon L, Beynnon Bruce D, Schottel Patrick C. Biomechanical comparison of a novel tensioned cable construct versus tension band wiring for transverse patella fracture fixation.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2022. DOI: 10.1007/s00590-022-03291-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.