MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.acvd.2023.10.006

Медицинская библиотека · Archives of cardiovascular diseases · 2023

Safety and efficacy of the Amplatzer™ Trevisio™ intravascular delivery system: Post-approval study results.

Hascoet Sebastien, Baruteau Alban-Elouen, Jalal Zakaria, Demkow Marcin, de Winter Robbert, Gaio Gianpiero, Clerc Jean-Michel, Sabiniewicz Robert, Eberli Franz, Santoro Giuseppe, Dauphin Claire, Schubert Stephan, Smolka Grzegorz, Lutz Matthias, Moreno Raul, Pan Manuel, Gutierrez-Larraya Federico, Godart Francois, Carminati Mario, Ovaert Caroline, Batteux Clement, Guerin Patrice, Thambo Jean-Benoit, Ewert Peter

Наблюдательное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The Amplatzer™ Trevisio™ Intravascular Delivery System (Trevisio DS; Abbott Laboratories, Chicago, IL, USA) facilitates the delivery of Amplatzer™ Occluders and features an ultraflexible tip, which improves assessment of occluder position before release.

AIMS: To assess the safety and efficacy of the Trevisio DS for transcatheter closure of patent foramen ovale and atrial septal defect.

METHODS

The Amplatzer™ Trevisio™ Intravascular Delivery System Post-Approval Study was a prospective, postmarket, single-arm, multicentre, observational study of the Trevisio DS. Enrolled patients were indicated for transcatheter closure of patent foramen ovale or atrial septal defect. In all procedures, the Trevisio DS was used to deliver Amplatzer™ Occluders. Technical success was defined as successful deployment and release of at least one occluder. Device- or procedure-related serious adverse events were tracked until discharge or day 7, whichever occurred earlier.

RESULTS

The study enrolled 144 patients with patent foramen ovale and 107 patients with atrial septal defect at 22 European sites; 53 patients with atrial septal defect (49.6%) were aged<18years. The rate of technical success was 98.4% (97.2% for atrial septal defect, 99.3% for patent foramen ovale). There was one serious adverse event (0.4%), an acute periprocedural device embolization that occurred after occluder release in a patient with atrial septal defect; the device was retrieved percutaneously. This was determined by the implanter to be unrelated to the performance of the Trevisio DS.

CONCLUSIONS

The Trevisio DS exhibited a high rate of technical success and an excellent safety profile during transcatheter closure of patent foramen ovale and atrial septal defect.

Для цитирования: Hascoet Sebastien, Baruteau Alban-Elouen, Jalal Zakaria, Demkow Marcin, de Winter Robbert, Gaio Gianpiero, Clerc Jean-Michel, Sabiniewicz Robert, Eberli Franz, Santoro Giuseppe, Dauphin Claire, Schubert Stephan, Smolka Grzegorz, Lutz Matthias, Moreno Raul, Pan Manuel, Gutierrez-Larraya Federico, Godart Francois, Carminati Mario, Ovaert Caroline, Batteux Clement, Guerin Patrice, Thambo Jean-Benoit, Ewert Peter. Safety and efficacy of the Amplatzer™ Trevisio™ intravascular delivery system: Post-approval study results.. Archives of cardiovascular diseases. 2023. DOI: 10.1016/j.acvd.2023.10.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.