Accelerometer-based vibration analysis and oxygenator thrombosis in venovenous ECMO: an experimental porcine model.

Antonsen Lars Prag, Landsverk Svein Aslak, Halvorsen Per Steinar, Thiara Amrit, Lilja Didrik, Nesaragi Naimahmed, Espinoza Andreas

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Oxygenator thrombosis is a potentially life-threatening complication during venovenous extracorporeal membrane oxygenation (VV ECMO). It can cause blood flow obstruction, impaired gas exchange, hematologic abnormalities, or sudden ECMO flow cessation. Early detection and timely circuit exchange is critical yet challenging. Acute clot formation necessitates immediate circuit replacement, while premature replacement risks unnecessary procedural harm and increased costs. No reliable method exists to detect early oxygenator thrombosis. Strategies include visual inspection, monitoring the pressure difference across the oxygenator (ΔPoxy), gas exchange evaluation, and blood tests. In the present animal study, we aimed to evaluate the feasibility of accelerometer-based vibration analysis as a real-time and non-invasive method for detecting oxygenator thrombosis during VV ECMO. We hypothesized that accelerometer signals would change concurrently with or precede increases in ΔPoxy.

METHODS

The study was performed on anesthetized and mechanically ventilated pigs (n = 7) on VV ECMO. Hemodynamic parameters, ECMO circuit pressures, and signals from an accelerometer attached to the ECMO oxygenator were continuously recorded at different pump speeds, and after anticoagulation reversal to promote thrombosis within the ECMO oxygenator.

RESULTS

The primary finding of this study was a significant increase in the accelerometer signal's Root Mean Squared (RMSoxy) 15 min after anticoagulation reversal, with no rpm adjustment and without corresponding changes in ΔPoxy. Variations in RMSoxy associated with high ECMO pump speed and circuit flow were discernible from those observed following anticoagulation reversal.

CONCLUSION

The present animal study demonstrates the feasibility of accelerometer-based vibration analysis as a real-time and non-invasive method for detecting vibrations associated with reversal of anticoagulation and potential oxygenator thrombosis during VV ECMO.

Для цитирования: Antonsen Lars Prag, Landsverk Svein Aslak, Halvorsen Per Steinar, Thiara Amrit, Lilja Didrik, Nesaragi Naimahmed, Espinoza Andreas. Accelerometer-based vibration analysis and oxygenator thrombosis in venovenous ECMO: an experimental porcine model.. Intensive care medicine experimental. 2025. DOI: 10.1186/s40635-025-00763-7.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.