Laboratory Protocol and Pilot Results for Dynamic Interference Testing of Continuous Glucose Monitoring Sensors.
Аннотация
BACKGROUND
Testing the potential influence of interfering substances on the measurement performance of needle sensors for continuous glucose monitoring (CGM) is a challenging task. For proper function, the sensors need an almost stable fluidic environment. Previously published in vitro interference experiments were measuring under static concentration conditons. Our experimental setup allows for interference testing with dynamic changes of the interferent concentrations.
METHODS
We designed a macrofluidic test stand that is fueled by several high-pressure liquid chromatography (HPLC) pumps generating programmable glucose and/or interferent gradients in phosphate-buffered saline (PBS). After optimizing experimental parameters (channel dimensions, temperature, flow rates, gradient slopes, buffer, pH etc.), we validated the setup using Dexcom G6 (G6) and Freestyle Libre 2 (L2) sensors with/without interferents, and using YSI 2300 Stat plus as the reference glucose device at room temperature.
RESULTS
Both sensors tracked the programmed glucose changes. After calibration, G6 results closely matched glucose reference readings, while L2 routinely showed ~50% to 60% lower readings, most likely because of the factory-based calibration and temperature compensation. Gradients of maltose, acetaminophen, and xylose were employed to further validate the setup. As expected, both sensors were not affected by maltose. We confirmed previous findings regarding susceptibility of G6 readings to acetaminophen and L2 readings to xylose. Signals from both sensors are influenced by temperature in a linear fashion.
CONCLUSIONS
Our experimental in vitro setup and protocol may provide a useful method to dynamically test CGM sensors for interfering substances. This may help to improve the accuracy of future CGM sensor generations.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.