Leakage of isopropanol from port protectors used in neonatal care-Results from an in vitro study.

Hjalmarsson Louise Björkman, Hagberg Jessika, Schollin Jens, Ohlin Andreas

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

To decrease contamination of needleless catheter hubs, caps or port protectors impregnated with isopropanol (IPA) have been developed and shown to be superior to other disinfection methods. The safety of the caps has been questioned, as they can be associated with alcohol leakage across the hub membrane.

OBJECTIVES

We evaluated the use of IPA caps and the scrub-the-hub method from the safety standpoint of possible alcohol leakage across the hub membrane.

METHODS

Circuits imitating an intravenous line were constructed. Circuits with an IPA cap were flushed with sodium chloride after the hub had been exposed to the cap for 1 hour, 24 hours, and 7 days. At the end of each period the fluid was collected and amounts of IPA in it were measured, using gas chromatography. Scrub circuits without IPA caps were also tested and ethanol from these was measured using the same method.

RESULTS

In this in vitro study, IPA was detected in all samples from cap circuits, and ethanol was detected from all scrub circuits. Leakage increased over time in IPA circuits. After 24 hours and 7 days of exposure, the first injection resulted in higher amounts of IPA; thereafter, the levels decreased. The amounts of ethanol measured from the scrub circuits were low.

CONCLUSIONS

IPA caps can cause leakage of alcohol across the hub membrane. Leakage increased over time, and a 30 sec drying time was not sufficient to solve the problem. Scrub-the-hub seems safe to use with regard to alcohol leakage.

Для цитирования: Hjalmarsson Louise Björkman, Hagberg Jessika, Schollin Jens, Ohlin Andreas. Leakage of isopropanol from port protectors used in neonatal care-Results from an in vitro study.. PloS one. 2020. DOI: 10.1371/journal.pone.0235593.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.