DOI 10.1016/j.jpeds.2020.08.079

Медицинская библиотека · The Journal of pediatrics · 2020

Decision Accuracy and Safety of Transcutaneous Bilirubin Screening at Intermountain Healthcare.

Konana Olive S, Bahr Timothy M, Strike Holly R, Coleman Jennifer, Snow Gregory L, Christensen Robert D

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To quantify the risk that transcutaneous bilirubin (TcB) screening would fail to recommend phototherapy for a neonate who would have qualified for it if total serum bilirubin (TSB) screening were used.

STUDY

DESIGN

We conducted a quality improvement project where simultaneous TcB and TSB were obtained on neonates ≥35 weeks of gestation during birth hospitalizations in our hospital system. Using our Utah bilirubin management algorithm, we quantified the risk that TcB screening would fail to identify the need for a confirmatory TSB when TSB screening alone would have revealed that phototherapy was indicated.

RESULTS

In 3 hospitals, we obtained 727 paired TcB/TSB measurements. Two instances utilized a blood gas radiometer for TSB, and 725 utilized the clinical laboratory-based TSB method. One of the 727 instances had a TcB indicating NO PHOTOTHERAPY, when the simultaneous TSB indicated PHOTOTHERAPY NEEDED. The TSB from that instance was 1 of the 2 from the blood gas radiometer. We estimate the risk of such an error occurring is 1.4 per 1000 TcB measurements (95% CI 0.03-7.6 per 1000). When only the laboratory TSB is used, we estimate the risk of such an error occurring to be 0 per 1000 TcB measurements (95% CI 0.0-5.1 per 1000).

CONCLUSIONS

Using TcB for screening at the birth hospital can identify those qualifying for phototherapy, using the Utah guidelines, with 1 of 727 neonates with a blood gas bilirubin and none of 725 with a laboratory-based analysis misidentified as not needing phototherapy when by TSB they did.

Для цитирования: Konana Olive S, Bahr Timothy M, Strike Holly R, Coleman Jennifer, Snow Gregory L, Christensen Robert D. Decision Accuracy and Safety of Transcutaneous Bilirubin Screening at Intermountain Healthcare.. The Journal of pediatrics. 2020. DOI: 10.1016/j.jpeds.2020.08.079.