DOI 10.1016/j.jvir.2010.04.020

Медицинская библиотека · Journal of vascular and interventional radiology : JVIR · 2010

Prospective randomized comparative evaluation of proximal valve polyurethane and distal valve silicone peripherally inserted central catheters.

Ong Cheng K, Venkatesh Sudhakar K, Lau Gabriel B, Wang Shih C

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

PURPOSE: The objective of this study was to evaluate and compare the relative durability and complications between the proximal valve polyurethane and distal valve silicone peripherally inserted central catheters (PICCs).

METHODS

Institutional review board approval was obtained. A total of 326 patients (mean age, 50.4 years) was assigned randomly to receive either a proximal valve polyurethane PICC (n = 198) or a distal valve silicone Groshong PICC (n = 194). All PICCs were inserted under radiologic guidance by interventional radiologists. Follow-up data were collected until catheter removal because of complications or treatment completion.

RESULTS

The mean catheter dwell time was 25.6 days (range, 1-245 days). Complications were encountered in 26.8% and 47.9% of the proximal valve polyurethane PICCs and distal valve silicone PICCs, respectively (P < .001). Significantly higher incidences of phlebitis (23.2% versus 11.6%, P = .003) and catheter-related infection (6.2% versus 2%, P = 0.043) were noted in the distal valve silicone PICCs. No significant differences in the incidence of catheter occlusion, fracture, or dislodgement were found. Multivariate logistic regression analysis showed a higher complication rate in the distal valve silicone PICCs corrected for patients' age, sex, underlying morbidity, indication, peripheral vein accessed, arm used, catheter tip placement, and the number of venepunctures attempted.

CONCLUSIONS

Proximal valve polyurethane PICCs were more durable than distal valve silicone PICCs, which were associated with a higher incidence of phlebitis and infection, probably related to the materials of the catheters and the designs and placements of the catheter valves.

Для цитирования: Ong Cheng K, Venkatesh Sudhakar K, Lau Gabriel B, Wang Shih C. Prospective randomized comparative evaluation of proximal valve polyurethane and distal valve silicone peripherally inserted central catheters.. Journal of vascular and interventional radiology : JVIR. 2010. DOI: 10.1016/j.jvir.2010.04.020.