Validity and reliability of electroacoustic probe for diagnosis of developmental dysplasia of the hip.

Padilla-Raygoza Nicolas, Olvera-Villanueva Georgina, Del Carmen Delgado-Sandoval Silvia, Cordova-Fraga Teodoro, Sosa-Aquino Modesto Antonio, Beltran-Campos Vicente

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Sound transmission is used in the diagnosis of hip dysplasia since the end of the 80's. Aim of this study is to quantify the validity and reliability of electroacoustic probe for the diagnosis of hip dysplasia in neonates.

METHODS

Diagnostic study included neonates aged 4-28 days, whose parents signed an informed consent. The probe was used three times for comparative sound transmission and with extension/flexion; hip ultrasound was performed with Graf technique as gold standard. Kappa was determined for intraobserver and interobserver reliability; validity was calculated with sensitivity, specificity, and predictive values.

RESULTS

100 neonates were included. For the comparative sound transmission, 0.80 and 0.81 Kappa were obtained for the intraobserver and interobserver respectively; with extension/flexion, Kappa 0.98 and 0.95 were obtained for the intraobserver and interobserver respectively. With comparative sound transmission, 44.8%, 97.7%, 76.5% and 91.3% for sensitivity, specificity, positive and negative predictive values, respectively; with extension/flexion test, the sensitivity, specificity, positive and negative predictive values: 82.8%, 99.4%, 96.0%, and 97.1%, respectively.

CONCLUSION

The electroacoustic probe is moderate valid and reliable for the diagnosis of developmental dysplasia of the hip.

TRIAL REGISTRATION: Open Science framework https://osf.io/kpf5s/?view_only=0a9682c6w1c842ad8e1d9a66e8dcf038.

Для цитирования: Padilla-Raygoza Nicolas, Olvera-Villanueva Georgina, Del Carmen Delgado-Sandoval Silvia, Cordova-Fraga Teodoro, Sosa-Aquino Modesto Antonio, Beltran-Campos Vicente. Validity and reliability of electroacoustic probe for diagnosis of developmental dysplasia of the hip.. BMC pediatrics. 2017. DOI: 10.1186/s12887-017-0903-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.