Spurious Hyperchloremia in the Presence of Elevated Plasma Salicylate: A Cohort Study.
Аннотация
BACKGROUND
Acute metabolic acidosis is rarely associated with a reduced or negative anion gap (AG), but several case reports have described such an abnormality occurring in the setting of acute salicylate intoxication. The underlying cause of this phenomenon is unclear.
METHODS
In this retrospective cohort study, we reviewed our institutional database to identify all patients admitted for salicylate intoxication at Mayo Clinic (Rochester, MN, USA) from January 2010 through December 2012. Serum chloride was measured with the Cobas INTEGRA 400 plus electrode (expedited laboratory test) or Cobas 6000 (routine laboratory test). We compared blood chloride levels measured by the 2 devices in the presence of positive blood salicylate level.
RESULTS
Twelve adult patients with salicylate levels >20 mg/dL had markedly elevated chloride concentrations. The median (interquartile range) chloride level at admission was 120 (107-145) mmol/L on their initial laboratory studies, resulting in reduced or even negative AGs. None of the patients had bromide toxicity, nor did they have any other identifiable cause of hyperchloremia or decreased AG. Further testing of the same blood samples with an alternative measurement system (Roche Cobas 6000) yielded normal chloride values, indicating that falsely elevated chloride values with the initial testing led to the diminished or negative AG values.
CONCLUSION
Circulating levels of salicylate can interfere with chloride measured by using routine techniques, resulting in spurious hyperchloremia outcomes and erroneous AG values. In patients with acute metabolic acidosis and abnormally reduced or negative AG, salicylate interference with chloride measurement should be suspected.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.