Continuous Intraoperative Neuromonitoring Study Using Pigs for the Prevention of Mechanical Recurrent Laryngeal Nerve Injury in Esophageal Surgery.
Аннотация
PURPOSES: During esophageal surgery, clamping injury and injury associated with the use of energy devices are common mechanisms underlying intraoperative recurrent laryngeal nerve (RLN) damage. Recently, intraoperative neuromonitoring (IONM) has been applied to prevent RLN injury. This study was aimed at investigating the changes in the EMG signals associated with clamping injury of the RLN caused by picking up of the nerve with tweezers in domestic pigs.
METHODS
Six domestic pigs (12 RLNs) underwent continuous IONM (CIONM) by our original automated periodic vagal nerve stimulation method.
RESULTS
Our system can be used safely and accurately. The signals showed a decrease of the amplitude when the RLN was picked up and closed slowly by the double-action Maryland with jaw covers. If the clamp was released before the signal amplitude decreased to 50% of the baseline, the signal showed gradual recovery to the baseline in 12 ± 3 minutes.
CONCLUSION
Although there were limitations in our study using domestic pig, including the small sample size, our results are expected to contribute to a decrease in the incidence of RLN damage during esophageal surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.