MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/tme.70103

Медицинская библиотека · Transfusion medicine (Oxford, England) · 2026

Simple image analysis for reliability control of column agglutination test in pretransfusion testing.

Pedersen Martin Töffner, Rajaguru Tharisa, Ashong Albert, Arbi Ramis, Hildebrandt Niko, Skovsted Thor

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

A novel meniscus-based analysis provides a rapid method to identify analytical errors related to unrecognised pipetting errors.

BACKGROUND

The indirect antihuman globulin test (IAT) is part of pretransfusion testing, enabling detection of erythrocyte alloantibodies. Failure to detect such antibodies may result in transfusion complications. In column agglutination technique (CAT)-based IAT, correct separation of the different liquids by an air gap during incubation is essential. Penetration of this air gap due to pipetting errors can cause neutralisation of AHG, resulting in false-negative or weak test results.

MATERIALS AND

METHODS

We investigated the fluid dynamics of CAT-based IAT and developed a simple image-based quality control approach to detect air gap penetration. Using routinely generated analyser images, the meniscus contact angle was quantified with the software ImageJ in 132 simulated crossmatch tests.

RESULTS

Receiver operating characteristic analysis demonstrated high discriminatory performance (AUC 0.935). Additional experiments using bovine serum albumin solutions confirmed that increased protein concentration significantly reduces meniscus contact angle, supporting the proposed mechanism of error prediction.

CONCLUSION

While further validation across platforms and laboratories is required, this approach has the potential to enhance test reliability, support troubleshooting, and improve patient safety in routine IAT CAT workflows.

Для цитирования: Pedersen Martin Töffner, Rajaguru Tharisa, Ashong Albert, Arbi Ramis, Hildebrandt Niko, Skovsted Thor. Simple image analysis for reliability control of column agglutination test in pretransfusion testing.. Transfusion medicine (Oxford, England). 2026. DOI: 10.1111/tme.70103.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.