DOI 10.1093/tropej/fmaa026

Медицинская библиотека · Journal of tropical pediatrics · 2020

Accuracy of Bilistick (a Point-of-Care Device) to Detect Neonatal Hyperbilirubinemia.

Kamineni Bhargavi, Tanniru Anusha, Vardhelli Venkateshwarlu, Sharma Deepak, Pawale Dinesh, Kulkarni Dattatray, Muppidi Pranitha Reddy, Deshabhotla Saikiran, Murki Srinivas

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Early diagnosis and appropriate management of neonatal jaundice is crucial in avoiding severe hyperbilirubinemia and brain injury. A low-cost, minimally invasive, point-of-care (PoC) tool for total bilirubin (TB) estimation which can be useful across all ranges of bilirubin values and all settings is the need of the hour.

OBJECTIVE

To assess the accuracy of Bilistick system, a PoC device, for measurement of TB in comparison with estimation by spectrophotometry.

DESIGN/

METHODS

In this cross-sectional clinical study, in infants who required TB estimation, blood samples in 25-µl sample transfer pipettes were collected at the same time from venous blood obtained for laboratory bilirubin estimation. The accuracy of Bilistick in estimating TB within ±2 mg/dl of bilirubin estimation by spectrophotometry was the primary outcome.

RESULTS

Among the enrolled infants, 198 infants were eligible for study analysis with the mean gestation of 36 ± 2.3 weeks and the mean birth weight of 2368 ± 623 g. The median age at enrollment was 68.5 h (interquartile range: 48-92). Bilistick was accurate only in 54.5% infants in measuring TB within ±2 mg/dl difference of TB measured by spectrophotometry. There was a moderate degree of correlation between the two methods (r = 0.457; 95% CI: 0.339-0.561, p value < 0.001). Bland-Altman analysis showed a mean difference of 0.5 mg/dl (SD ± 4.4) with limits of agreement between -8.2 and +9.1 mg/dl.

CONCLUSION

Bilistick as a PoC device is not accurate to estimate TB within the clinically acceptable difference (±2 mg/dl) of TB estimation by spectrophotometry and needs further improvement to make it more accurate.

Для цитирования: Kamineni Bhargavi, Tanniru Anusha, Vardhelli Venkateshwarlu, Sharma Deepak, Pawale Dinesh, Kulkarni Dattatray, Muppidi Pranitha Reddy, Deshabhotla Saikiran, Murki Srinivas. Accuracy of Bilistick (a Point-of-Care Device) to Detect Neonatal Hyperbilirubinemia.. Journal of tropical pediatrics. 2020. DOI: 10.1093/tropej/fmaa026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.