MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ijgo.13475

Медицинская библиотека · International journal of gynaecology and obstetrics: the official organ of the International Federation of Gynaecology and Obstetrics · 2020

Reliability of ultrasound findings acquired with handheld apparatuses to inform urgent obstetric diagnosis in a high-volume resource-limited setting.

Kodaira Yuichi, Pisani Luigi, Boyle Sonia, Olumide Salawu, Orsi Michele, Adeniji Adetunji O, Pisani Enzo, Zanette Marianna, Putoto Giovanni, Koroma Michael M

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To evaluate the reliability of obstetric handheld smartphone-based point-of-care ultrasound (POCUS) in a resource-limited high-volume setting.

METHODS

A single-center prospective observational study among women requiring urgent diagnosis and admitted to a maternity referral hospital in Sierra Leone from March to April 2019. Pre-specified ultrasound findings were obtained with a handheld POCUS device; a comprehensive ultrasound examination was then performed by an experienced operator using conventional full-feature apparatus. Agreement was assessed by diagnostic accuracy and Cohen κ-statistics.

RESULTS

Overall, there were 307 participants. The mean aggregated diagnostic accuracy was 95.5% (κ-statistic, 0.90; 95% confidence interval [CI], 0.89-0.93; P < 0.001). Highest accuracy was reported for detecting free fluid collection in the abdominal cavity (100%; κ-statistic, 1.00; 95% CI, 1.00-1.00; P < 0.001). Ultrasound findings obtained with the handheld device for intrauterine pregnancy, fetal heartbeat, cephalic presentation, multifetal pregnancy, and assessment of gestational age based on bi-parietal diameter were highly reliable (agreement, >90%; κ-statistic, >0.80). Detection of low-lying placenta or placenta previa was the least reliable (κ-statistic, 0.53; 95% CI, 0.13-0.93; P < 0.001).

CONCLUSION

Handheld POCUS findings were found to be reliable for detecting pre-specified urgent obstetric findings in a high-volume resource-limited referral hospital.

Для цитирования: Kodaira Yuichi, Pisani Luigi, Boyle Sonia, Olumide Salawu, Orsi Michele, Adeniji Adetunji O, Pisani Enzo, Zanette Marianna, Putoto Giovanni, Koroma Michael M. Reliability of ultrasound findings acquired with handheld apparatuses to inform urgent obstetric diagnosis in a high-volume resource-limited setting.. International journal of gynaecology and obstetrics: the official organ of the International Federation of Gynaecology and Obstetrics. 2020. DOI: 10.1002/ijgo.13475.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.