MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/bjr/tqae006

Медицинская библиотека · The British journal of radiology · 2024

Sixty-four-fold data reduction of chest radiographs using a super-resolution convolutional neural network.

Nam Ju Gang, Kang Seung Kwan, Choi Hyewon, Hong Wonju, Park Jongsoo, Goo Jin Mo, Lee Jae Sung, Park Chang Min

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To develop and validate a super-resolution (SR) algorithm generating clinically feasible chest radiographs from 64-fold reduced data.

METHODS

An SR convolutional neural network was trained to produce original-resolution images (output) from 64-fold reduced images (input) using 128 × 128 patches (n = 127 030). For validation, 112 radiographs-including those with pneumothorax (n = 17), nodules (n = 20), consolidations (n = 18), and ground-glass opacity (GGO; n = 16)-were collected. Three image sets were prepared: the original images and those reconstructed using SR and conventional linear interpolation (LI) using 64-fold reduced data. The mean-squared error (MSE) was calculated to measure similarity between the reconstructed and original images, and image noise was quantified. Three thoracic radiologists evaluated the quality of each image and decided whether any abnormalities were present.

RESULTS

The SR-images were more similar to the original images than the LI-reconstructed images (MSE: 9269 ± 1015 vs. 9429 ± 1057; P = .02). The SR-images showed lower measured noise and scored better noise level by three radiologists than both original and LI-reconstructed images (Ps < .01). The radiologists' pooled sensitivity with the SR-reconstructed images was not significantly different compared with the original images for detecting pneumothorax (SR vs. original, 90.2% [46/51] vs. 96.1% [49/51]; P = .19), nodule (90.0% [54/60] vs. 85.0% [51/60]; P = .26), consolidation (100% [54/54] vs. 96.3% [52/54]; P = .50), and GGO (91.7% [44/48] vs. 95.8% [46/48]; P = .69).

CONCLUSIONS

SR-reconstructed chest radiographs using 64-fold reduced data showed a lower noise level than the original images, with equivalent sensitivity for detecting major abnormalities.

ADVANCES IN KNOWLEDGE: This is the first study applying super-resolution in data reduction of chest radiographs.

Для цитирования: Nam Ju Gang, Kang Seung Kwan, Choi Hyewon, Hong Wonju, Park Jongsoo, Goo Jin Mo, Lee Jae Sung, Park Chang Min. Sixty-four-fold data reduction of chest radiographs using a super-resolution convolutional neural network.. The British journal of radiology. 2024. DOI: 10.1093/bjr/tqae006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.