Evaluation of two-dimensional cytometric lamellar body counts on the ADVIA 120 hematology system for estimation of fetal lung maturation.

Chapman John F, Ashwood Edward R, Feld Ronald, Wu Alan H B

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

We sought to determine if the ADVIA 120 hematology system (Bayer HealthCare LLC, Diagnostics Division, Tarrytown, NY), which employs a unique two-dimensional cytometric analysis approach for platelet (PLT) counting, might also be useful for estimation of LB lamellar body counts in amniotic fluid.

METHODS

ADVIA 120 LB counts were performed on 217 amniotic fluid specimens, 88 of which were obtained within 72 h of infant delivery. The ADVIA 120 LB count (ADVIA LB count) results were compared with results from other FLM tests, including the TDx-FLM II (FLM II) test (Abbott Diagnostics, Abbott Park, IL) and the A(650) estimate of lamellar body density using the refractive-index-matched anomalous diffraction (RIMAD) technique.

RESULTS

Use of an ADVIA LB count referent value of > or =35,400/microl to indicate maturity, yielded sensitivity, specificity, PV(RDS) and PV(maturity) of 100%, 67.6%, 36.8% and 100%, respectively, in our study population (prevalence of RDS=15.9%). The clinical performance of the ADVIA LB count assay was compared to that of the FLM II and RIMAD assays by predictive value and receiver operating characteristic (ROC) curve analysis.

CONCLUSIONS

These data suggest that the ADVIA 120-derived LB counts on amniotic fluid can be useful in predicting fetal lung maturity.

Для цитирования: Chapman John F, Ashwood Edward R, Feld Ronald, Wu Alan H B. Evaluation of two-dimensional cytometric lamellar body counts on the ADVIA 120 hematology system for estimation of fetal lung maturation.. Clinica chimica acta; international journal of clinical chemistry. 2004. DOI: 10.1016/j.cccn.2003.09.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.