New retrievable coil anchors: preliminary in vivo experiences in swine.
Аннотация
PURPOSE: To design and test retrievable coil anchors to improve the safety and efficacy of coil embolization.
METHODS
Fifty-two 0.038-inch homemade retrievable stainless steel coils were equipped with one of four different pre-shaped nitinol anchors and tested in 38 pigs. All coils with the anchor were completely retrieved and redeployed 3-18 times (median 7 times) prior to release. Types 1 and 2 anchored coils were acutely deployed in the external iliac arteries (n = 10 each), and chronically tested (1 week) in the common carotid arteries (n = 6 each). Larger type 1 (n = 4), type 3 (n = 6), and type 4 (n = 4) anchored coils were acutely deployed in the abdominal aorta. The largest type 1 anchors (n = 6) were acutely tested in the inferior vena cava.
RESULTS
All anchored coils were successfully retrieved and repositioned several times. All but two coils formed a compact plug and there was no coil migration except with two mechanically defective type 3 anchors.
CONCLUSION
The use of retrievable anchors allowed the coils to be retrieved and repositioned, prevented coil migration, and enabled compact coil configuration.