Electrical impedance tomography: a method for monitoring regional lung aeration and tidal volume distribution?

Frerichs Inéz, Dargaville Peter A, Dudykevych Taras, Rimensberger Peter C

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To demonstrate the monitoring capacity of modern electrical impedance tomography (EIT) as an indicator of regional lung aeration and tidal volume distribution.

DESIGN AND

SETTING

Short-term ventilation experiment in an animal research laboratory.

PATIENTS AND

PARTICIPANTS

One newborn piglet (body weight: 2 kg).

INTERVENTIONS: Surfactant depletion by repeated bronchoalveolar lavage, surfactant administration.

MEASUREMENTS AND

RESULTS

EIT scanning was performed at an acquisition rate of 13 images/s during two ventilatory manoeuvres performed before and after surfactant administration. During the scanning periods of 120 s the piglet was ventilated with a tidal volume of 10 ml/kg at positive end-expiratory pressures (PEEP) in the range of 0-30 cmH(2)O, increasing and decreasing in 5 cmH(2)O steps. Local changes in aeration and ventilation with PEEP were visualised by EIT scans showing the regional shifts in end-expiratory lung volume and distribution of tidal volume, respectively. In selected regions of interest EIT clearly identified the changes in local aeration and tidal volume distribution over time and after surfactant treatment as well as the differences between stepwise inflation and deflation.

CONCLUSIONS

Our data indicate that modern EIT devices provide an assessment of regional lung aeration and tidal volume and allow evaluation of immediate effects of a change in ventilation or other therapeutic intervention. Future use of EIT in a clinical setting is expected to optimise the selection of appropriate ventilation strategies.

Для цитирования: Frerichs Inéz, Dargaville Peter A, Dudykevych Taras, Rimensberger Peter C. Electrical impedance tomography: a method for monitoring regional lung aeration and tidal volume distribution?. Intensive care medicine. 2003. DOI: 10.1007/s00134-003-2029-z.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.