Late presentation of CardioSEAL PFO closure device fracture and thrombus formation three years after device implantation.
Аннотация
Transcatheter closure of patent foramen ovale (PFO) is a promising alternative to surgical closure or anticoagulation therapy in patients with a history of cryptogenic stroke. Although a number of transcatheter closure devices have been developed and used outside of the United States, the two commonly used devices in the United States are the StarFLEX CardioSEAL and the Amplatzer Occluder. Several complications, both early and late, associated with the devices have been reported in 6-11% of patients. The rates of complication vary with the different devices. Complications include device-related thrombi, embolization, device malposition, fracture, air embolism, arrhythmias, pericardial effusion, need for surgical intervention and death. As reported in the literature, the incidence of longterm complications beyond 1 year is rare. We report a case of a very late presentation of CardioSEAL device arm fracture associated with left atrial thrombus 3 years after device implantation.