DOI 10.1016/j.ajem.2010.05.013

Медицинская библиотека · The American journal of emergency medicine · 2010

A novel hands-free carotid ultrasound detects low-flow cardiac output in a swine model of pulseless electrical activity arrest.

Larabee Todd M, Little Charles M, Raju Balasundar I, Cohen-Solal Eric, Erkamp Ramon, Wuthrich Scott, Petruzzello John, Nakagawa Michael, Ayati Shervin

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine if a hands-free, noninvasive Doppler ultrasound device can reliably detect low-flow cardiac output by measuring carotid artery blood flow velocities. We compared the ability of observers to detect carotid artery flow velocity differences between pseudo-pulseless electrical activity (PEA) and true-PEA cardiac arrest.

METHODS

Five swine were instrumented with aortic (Ao) and right atrial pressure-transducing catheters. The Doppler ultrasound device was adhered to the neck over the carotid artery. Continuous electrocardiogram, pressure readings, and Doppler signal were recorded. Each swine underwent multiple episodes of fibrillation and resuscitation. Episodes of true-PEA and pseudo-PEA were retrospectively identified from all resuscitation attempts by examination of electrocardiogram and Ao waveforms. The sensitivity and specificity of the device to detect pseudo-PEA was obtained using observers blinded to Ao waveform recordings.

RESULTS

There was good interobserver reliability related to identification of pseudo- and true-PEA (κ = 0.873). The observers blinded to Ao waveform recordings agreed on 8 of the 9 episodes of pseudo-PEA, whereas 4 false positives of 26 true-PEA events were reported (sensitivity, 0.89; specificity, 0.85). The Doppler device was able to detect carotid flow velocity over a wide range of Ao blood pressures.

CONCLUSIONS

This hands-free, noninvasive Doppler ultrasound device can reliably differentiate pseudo-PEA from true-PEA during resuscitation from cardiac arrest, detecting pressure gradient changes of less than 5 mm Hg through to normotension. This device distinguishes conditions of no cardiac output from low cardiac output and may have applications for use during resuscitation from various etiologies of arrest and shock.

Для цитирования: Larabee Todd M, Little Charles M, Raju Balasundar I, Cohen-Solal Eric, Erkamp Ramon, Wuthrich Scott, Petruzzello John, Nakagawa Michael, Ayati Shervin. A novel hands-free carotid ultrasound detects low-flow cardiac output in a swine model of pulseless electrical activity arrest.. The American journal of emergency medicine. 2010. DOI: 10.1016/j.ajem.2010.05.013.