Monopolar biphasic focal pulsed field ablation directly at the atrioventricular junction and from within the noncoronary cusp: The PFA-CONDUCT study.

Linz Dominik, Hertel Julie Norup, Chaldoupi Sevasti-Maria, Gil-Ramirez Andreina, Sattler Stefan M, Linz Benedikt, Farnir Florent I P, Saljic Arnela, Mohaissen Tasnim, Hermans Ben J M, Ruwald Martin H, Luermans Justin G L M, Keelani Ahmad, Geller J Christoph, Raffa Santi, Hansen Jim, Jespersen Thomas, Nissen Sarah Dalgas

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The efficacy and safety of monopolar biphasic focal pulsed field ablation (F-PFA) in close proximity to the atrioventricular (AV) node remain unknown.

OBJECTIVE

This study aimed to describe the effect of direct or indirect F-PFA application at the AV junction.

METHODS

In pigs, F-PFA (Centauri, CardioFocus) was applied directly at the AV junction (n = 3) and indirectly from within the aortic noncoronary cusp (NCC) (n = 5), followed by macroscopic gross and histologic analysis. In 5 patients planned for a pace-and-ablate strategy, F-PFA was applied at the AV junction. In 1 patient with recurrent para-Hisian atrial tachycardia and previously failed ablation, F-PFA was applied from within the NCC.

RESULTS

In pigs, direct F-PFA applications at the AV junction were associated with junctional beats and resulted in complete AV block after 2-3 F-PFA applications. Tissue histology revealed a lesion depth of approximately 8 mm with minor hemorrhage and leukocyte infiltration. F-PFA from within the NCC resulted in a transient increase in the AV node's Wenckebach cycle length and 1 pig had transient AV block, which recovered within 10 minutes. In 5 patients undergoing pace-and-ablate strategy, F-PFA applications at the AV junction resulted in sustained AV block. In 1 patient, application of F-PFA from within the NCC resulted in termination and noninducibility of a recurrent incessant para-Hisian atrial tachycardia without AV block. There were no immediate or long-term complications.

CONCLUSION

F-PFA directly at the AV junction results in nonreversible AV block. F-PFA from within the NCC may represent a promising strategy for treating para-Hisian arrhythmias while preserving AV conduction.

Для цитирования: Linz Dominik, Hertel Julie Norup, Chaldoupi Sevasti-Maria, Gil-Ramirez Andreina, Sattler Stefan M, Linz Benedikt, Farnir Florent I P, Saljic Arnela, Mohaissen Tasnim, Hermans Ben J M, Ruwald Martin H, Luermans Justin G L M, Keelani Ahmad, Geller J Christoph, Raffa Santi, Hansen Jim, Jespersen Thomas, Nissen Sarah Dalgas. Monopolar biphasic focal pulsed field ablation directly at the atrioventricular junction and from within the noncoronary cusp: The PFA-CONDUCT study.. Heart rhythm. 2026. DOI: 10.1016/j.hrthm.2026.01.012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.