Broadband near-infrared spectroscopy can detect cyanide-induced cytochrome aa3 inhibition in rats: a proof of concept study.

Thiele Robert H, Ikeda Keita, Wang Yunzhen, Bartz Raquel R, Zuo Zhiyi

СтатьяПеревод в очереди

Аннотация

PURPOSE: Clinically available near-infrared spectroscopy (NIRS) devices use two to five wavelengths of light to measure the relative amounts of oxyhemoglobin and deoxyhemoglobin in tissue to determine tissue hemoglobin oxygen saturation (StO2). In addition to StO2, broadband NIRS devices (using hundreds of wavelengths of light) may be able to measure the oxidation state of mitochondrial cytochrome aa3 (Cytox) which reflects the subcellular energetic state. We hypothesize that broadband NIRS devices can measure Cytox independent of changes in hemoglobin saturation.

METHODS

In this prospective non-randomized study, 20 male Sprague-Dawley rats (300 g) were anesthetized with isoflurane, tracheally intubated, and ventilated with 100% O2 containing 2% isoflurane. They were subsequently instrumented with a broadband NIRS device that used a halogen light source coupled to an emitting fibreoptic cable. Three receiving fibreoptic cables were utilized; one analyzed the light source and the other two were directed at the base of the skull. Each receiving fibre was connected to a spectrometer to measure light intensity. Sodium cyanide (NaCN) 5 mg·kg-1 iv was injected in order to produce cytochrome aa3 reduction. Two to three minutes after injection, oxygen was eliminated and 100% nitrogen (i.e., anoxia) was used for ventilation in order to induce a reduction in both cytochrome aa3 and hemoglobin desaturation. Changes in the cytochrome oxidation state and hemoglobin oxygenation were calculated using a broadband algorithm and compared before and after both the NaCN and anoxia interventions.

RESULTS

The NaCN injection resulted in a decrease in median [interquartile range (IQR)] deoxyhemoglobin (-0.014 [-0.29 to -0.005] arbitrary units [AU]; P < 0.001), an increase in oxyhemoglobin (0.013 [-0.011 to 0.031] AU; P < 0.001), and a reduction in cytochrome aa3 (-0.015 [-0.020 to -0.011] AU; P < 0.001). Anoxia resulted in an increase in median [IQR] deoxyhemoglobin (0.13 [0.11 to 0.18] AU; P < 0.001), a decrease in oxyhemoglobin (-0.17 [-0.22 to -0.15] AU; P < 0.001), and a reduction in cytochrome aa3 (-0.04 [-0.06 to -0.03] AU; P < 0.001).

CONCLUSION

Broadband NIRS can effectively measure the directionality of changes in both Cytox and StO2 by uncoupling the cytochrome and hemoglobin signals through inhibition of the electron transport chain and anoxia.

Для цитирования: Thiele Robert H, Ikeda Keita, Wang Yunzhen, Bartz Raquel R, Zuo Zhiyi. Broadband near-infrared spectroscopy can detect cyanide-induced cytochrome aa3 inhibition in rats: a proof of concept study.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 2016. DOI: 10.1007/s12630-016-0803-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.