DOI 10.1016/j.joms.2011.02.111
Venous anastomoses using microvascular coupler in free flap head and neck reconstruction.
Аннотация
PURPOSE: This study is a retrospective review of the experience using the venous coupler for head and neck reconstruction over a 3-year period at the University of Maryland Medical Center, Department of Oral and Maxillofacial Surgery.
MATERIALS AND
METHODS
One hundred seventy-eight consecutive cases of microvascular free flaps between May 2007 and September 2010 were retrospectively reviewed. Data were collected by demographic information, flap type, recipient vessels, method of anastomosis, coupler size, coupler orientation, complications associated with coupler, and reconstruction results. Fisher exact test was used for statistical analysis.
RESULTS
There were 294 anastomotic coupler devices used in 173 flaps, with hand-sewn venous anastomoses performed in 5 patients. The overall flap success rate was 94.9% (169/178). Success rate among cases in which the coupler was used was 95.4% (8/173). Total coupler venous thrombosis rate was 4.0% (7/173), with a statistically significant difference (P < .05) in reference to the number of venous anastomoses performed: 58 cases had a single vein anastomosed, 5 cases developed thrombosis; while the 115 flaps with 2 venous anastomoses, only 2 cases had thrombosis.
CONCLUSIONS
The microvascular coupler is reliable for venous anastomosis in free flap head and neck reconstruction; dual-vein anastomoses appear to have better results than single-vein anastomoses. Flow coupler has a promising utility in monitoring buried flaps and flaps that are difficult to observe. The microvascular coupler deserves to be more commonly used in free flap head and neck reconstruction.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.