Analysis of adverse events with Essure hysteroscopic sterilization reported to the Manufacturer and User Facility Device Experience database.

Al-Safi Zain A, Shavell Valerie I, Hobson Deslyn T G, Berman Jay M, Diamond Michael P

СтатьяПеревод в очереди

Аннотация

STUDY

OBJECTIVE

The Manufacturer and User Facility Device Experience database may be useful for clinicians using a Food and Drug Administration-approved medical device to identify the occurrence of adverse events and complications. We sought to analyze and investigate reports associated with the Essure hysteroscopic sterilization system (Conceptus Inc., Mountain View, CA) using this database.

DESIGN

Retrospective review of the Manufacturer and User Facility Device Experience database for events related to Essure hysteroscopic sterilization from November 2002 to February 2012 (Canadian Task Force Classification III).

SETTING

Online retrospective review.

PATIENTS: Online reports of patients who underwent Essure tubal sterilization.

INTERVENTION: Essure tubal sterilization.

MEASUREMENTS AND MAIN

RESULTS

Four hundred fifty-seven adverse events were reported in the study period. Pain was the most frequently reported event (217 events [47.5%]) followed by delivery catheter malfunction (121 events [26.4%]). Poststerilization pregnancy was reported in 61 events (13.3%), of which 29 were ectopic pregnancies. Other reported events included perforation (90 events [19.7%]), abnormal bleeding (44 events [9.6%]), and microinsert malposition (33 events [7.2%]). The evaluation and management of these events resulted in an additional surgical procedure in 270 cases (59.1%), of which 44 were hysterectomies.

CONCLUSION

Sixty-one unintended poststerilization pregnancies were reported in the study period, of which 29 (47.5%) were ectopic gestations. Thus, ectopic pregnancy must be considered if a woman becomes pregnant after Essure hysteroscopic sterilization. Additionally, 44 women underwent hysterectomy after an adverse event reported to be associated with the use of the device.

Для цитирования: Al-Safi Zain A, Shavell Valerie I, Hobson Deslyn T G, Berman Jay M, Diamond Michael P. Analysis of adverse events with Essure hysteroscopic sterilization reported to the Manufacturer and User Facility Device Experience database.. Journal of minimally invasive gynecology. 2013. DOI: 10.1016/j.jmig.2013.05.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.