Does the type of suture technique affect the fluid-dynamic performance of bioprostheses implanted in small aortic roots? Results from an in vitro study.

Tasca Giordano, Vismara Riccardo, Fiore Gianfranco Beniamino, Romagnoni Claudia, Redaelli Alberto, Antona Carlo, Gamba Amando

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The in vivo hemodynamic performance of a bioprosthesis implanted in an aortic position is affected by the characteristics of the prosthesis and the sizing strategy adopted. Recently, it has been hypothesized that the type of suture used to implant the prosthesis might influence hemodynamics.

METHODS

Bioprostheses with labeled sizes of 19 mm and 21 mm were implanted in 2 groups of 5 porcine aortic roots, with native annuli of 19 mm and 21 mm, by means of 2 different suture techniques: simple interrupted and noneverting mattress with pledgets. The aortic roots were tested in an in vitro mock loop. The stroke volume imposed by the mock loop was set at 40 mL, and was increased by steps of 15 mL until a stroke volume of 100 mL was attained. Main fluid-dynamic parameters were analyzed.

RESULTS

At each level of stroke volume, ie, 40 mL, 55 mL, 70 mL, 85 mL, and 100 mL, the mean and peak pressure drops were significantly greater with the noneverting mattress suture with pledgets than with the simple interrupted suture. The effective orifice area behaved accordingly, being significantly smaller in the former case.

CONCLUSIONS

Our data show that the type of suture technique can influence bioprosthesis performance and that it is reasonable to assume that this is especially true in small annuli (≤ 21 mm). Thus, to optimize prosthesis performance and reduce the incidence of patient-prosthesis mismatch, the role of the suture technique should not be disregarded.

Для цитирования: Tasca Giordano, Vismara Riccardo, Fiore Gianfranco Beniamino, Romagnoni Claudia, Redaelli Alberto, Antona Carlo, Gamba Amando. Does the type of suture technique affect the fluid-dynamic performance of bioprostheses implanted in small aortic roots? Results from an in vitro study.. The Journal of thoracic and cardiovascular surgery. 2014. DOI: 10.1016/j.jtcvs.2014.10.096.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.