Continuous chest compressions with a simultaneous triggered ventilator in the Munich Emergency Medical Services: a case series.
Клинический случайСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Schaller Stefan J, Altmann Sonja, Unsworth Annalise, Schneider Gerhard, Bogner-Flatz Viktoria, Paul Thomas, Hoppmann Petra, Kanz Karl-Georg
Аннотация
Background: Mechanical chest compression devices are commonly used providing a constant force and frequency of chest compression during cardiopulmonary resuscitation. However, there are currently no recommendations on ventilation during cardiopulmonary resuscitation with a mechanical chest compression device using continuous mode. An effective method for ventilation in such scenarios might be a triggered oxygen-powered resuscitator. Methods: We report seven cardiopulmonary resuscitation cases from the Munich Emergency Medical Service where mechanical chest compression devices in continuous mode were used with an oxygen-powered resuscitator. In each case, the resuscitator (Oxylator®) was running in automatic mode delivering a breath during the decompression phase of the chest compressions at a frequency of 100 per minute. End-tidal carbon dioxide and pulse oximetry were measured. Additional data was collected from the resuscitation protocol of each patient. Results: End-tidal carbon dioxide was available in all cases while oxygen saturation only in four. Five patients had a return of spontaneous circulation. Based on the end-tidal carbon dioxide values of each of the cases, the resuscitator did not seem to cause hyperventilation and suggests that good-quality cardiopulmonary resuscitation was delivered. Conclusions: Continuous chest compressions using a mechanical chest compression device and simultaneous synchronized ventilation using an oxygen-powered resuscitator in an automatic triggering mode might be feasible during cardiopulmonary resuscitation.