DOI 10.1016/j.parkreldis.2023.105435
Acute stimulation with symmetric biphasic pulses induces less ataxia compared to cathodic pulses in DBS for essential tremor.
Аннотация
BACKGROUND
Symmetric biphasic pulses have been shown to acutely increase the therapeutic window of ventralis intermedius deep brain stimulation (Vim-DBS) for essential tremor (ET) compared to cathodic pulses. Acute supratherapeutic stimulation can induce ataxic side effects in Vim-DBS.
OBJECTIVE
To investigate the effect on tremor, ataxia and dysarthria of 3 h of biphasic stimulation in patients with DBS for ET.
METHODS
A randomized, doubled-blind, cross-over design was used to compare standard cathodic pulses with symmetric biphasic pulses (anode-first) during a 3-h period per pulse shape. During each 3-h period, all stimulation parameters were identical, except for the pulse shape. Tremor (Fahn-Tolosa-Marin Tremor Rating Scale), ataxia (International Cooperative Ataxia Rating Scale) and speech (acoustic and perceptual measures) were assessed hourly during the 3-h periods.
RESULTS
Twelve ET patients were included. During the 3-h stimulation period, tremor control was equivalent between the two pulse shapes. Biphasic pulses elicited significantly less ataxia than cathodic pulses (p = 0.006). Diadochokinesis rate of speech was better for the biphasic pulse (p = 0.048), but other measures for dysarthria were not significantly different between the pulses.
CONCLUSION
Symmetric biphasic pulses induce less ataxia than conventional pulses after 3 h of stimulation DBS in ET patients.