MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.resuscitation.2014.09.013

Медицинская библиотека · Resuscitation · 2014

Hemodynamic improvement of a LUCAS 2 automated device by addition of an impedance threshold device in a pig model of cardiac arrest.

Debaty Guillaume, Segal Nicolas, Matsuura Tim, Fahey Brian, Wayne Marvin, Mahoney Brian, Frascone Ralph, Lick Charles, Yannopoulos Demetris

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

The combination of the LUCAS 2 (L-CPR) automated CPR device and an impedance threshold device (ITD) has been widely implemented in the clinical field. This animal study tested the hypothesis that the addition of an ITD on L-CPR would enhance cerebral and coronary perfusion pressures.

METHODS

Ten female pigs (39.0 ± 2.0 kg) were sedated, intubated, anesthetized with isofluorane, and paralyzed with succinylcholine (93.3 μg/kg/min) to inhibit the potential confounding effect of gasping. After 4 min of untreated ventricular fibrillation, 4 min of L-CPR+an active ITD or L-CPR+a sham ITD was initiated and followed by another 4 min of the alternative method of CPR. Systolic blood pressure (SBP), diastolic blood pressure (DBP), diastolic right atrial pressure (RAP), intracranial pressure (ICP), airway pressure, and end tidal CO2 (ETCO2) were recorded continuously. Data expressed as mean mmHg ± SD.

RESULTS

Decompression phase airway pressure was significantly lower with L-CPR+active ITD versus L-CPR+sham ITD (-5.3 ± 2.2 vs. -0.5 ± 0.6; p<0.001). L-CPR+active ITD treatment resulted in significantly improved hemodynamics versus L-CPR+sham ITD: ETCO2, 35 ± 6 vs. 29 ± 7 (p=0.015); SBP, 99 ± 9 vs. 93 ± 15 (p=0.050); DBP, 24 ± 12 vs. 19 ± 15 (p=0.006); coronary perfusion pressure, 29 ± 8 vs. 26 ± 7 (p=0.004) and cerebral perfusion pressure, 24 ± 13 vs. 21 ± 12 (p=0.028).

CONCLUSIONS

In pigs undergoing L-CPR the addition of the active ITD significantly reduced intrathoracic pressure and increased vital organ perfusion pressures.

Для цитирования: Debaty Guillaume, Segal Nicolas, Matsuura Tim, Fahey Brian, Wayne Marvin, Mahoney Brian, Frascone Ralph, Lick Charles, Yannopoulos Demetris. Hemodynamic improvement of a LUCAS 2 automated device by addition of an impedance threshold device in a pig model of cardiac arrest.. Resuscitation. 2014. DOI: 10.1016/j.resuscitation.2014.09.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.