MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-0035-1566266

Медицинская библиотека · The Thoracic and cardiovascular surgeon reports · 2015

Long-Range Real Migration of Implantable Cardioverter Defibrillator Lead.

Keyser Andreas, Brodoefel Harald, Schmid Christof

СтатьяПеревод в очереди

Аннотация

The need for pacemaker and implantable cardioverter defibrillator (ICD) lead revisions and extractions is steadily increasing. Despite the lack of representative studies, the risk of lead extraction is frequently considered to be lower than leaving nonfunctional leads in situ. We report the case of a patient who was referred to our institution for exchange of a malfunctioning ICD lead. The diagnostic work-up revealed a long-segment transmural migration of the ICD lead at the site of the subclavian and innominate vein. Due to the unpredictable risk of vein perforation, we abandoned the extraction procedure.

Для цитирования: Keyser Andreas, Brodoefel Harald, Schmid Christof. Long-Range Real Migration of Implantable Cardioverter Defibrillator Lead.. The Thoracic and cardiovascular surgeon reports. 2015. DOI: 10.1055/s-0035-1566266.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.