Heat generation as an index of exhaustion of soda lime.
Аннотация
The wall temperatures of the absorptive chambers of a divided soda lime canister were measured in 70 patients to determine the relationship between the difference in temperature of the two chambers and CO2 passage through the first chamber. CO2 passage through the first chamber was detected when the temperature of the second chamber became equal to that of the first. A significant correlation (R = 0.94; P less than 0.001) was found between the magnitude of CO2 passage through the first chamber and the difference in temperature between the chambers. When the maximal absorptive capacity of soda lime was reached, the pH of the surface of soda lime granules was still too high to change the indicator color. Exhaustion of soda lime is more reliably recognized by measuring wall temperatures of the chambers than by observing color change of the soda lime granules.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.