Медицинская библиотека · Acta neurochirurgica · 2020

Paradoxical response of intracranial pressure to shunt valve setting adjustments.

D'Antona Linda, Craven Claudia Louise, Merchan Melida Andrea Jaime, Thompson Simon David, Bremner Fion, Thorne Lewis, Matharu Manjit Singh, Watkins Laurence Dale, Toma Ahmed Kassem

СтатьяНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

The hydrodynamics of cerebrospinal fluid shunts have been described in vitro; however, knowledge on the response of intracranial pressure (ICP) to valve settings adjustments in vivo is limited. This study describes the effect of adjusting the shunt valve setting on ICP in a cohort of patients with complex symptom management.

METHOD: Single-centre retrospective observational study. Patients who underwent ICP-guided valve setting adjustments during 24-h continuous ICP monitoring, between 2014 and 2019, were included. Patients with suspected shunt malfunction were excluded. Median night ICP before and after the valve adjustments were compared (Δ night ICP). The responses of ICP to valve adjustment were divided into 3 different groups as follows: expected, paradoxical and no response. The frequency of the paradoxical response and its potential predicting factors were investigated.

RESULTS

Fifty-one patients (37 females, 14 males, mean age 38 years) receiving 94 valve setting adjustments met the study inclusion criteria. Patients' underlying conditions were most commonly hydrocephalus (47%) or idiopathic intracranial hypertension (43%). The response of ICP to valve setting adjustments was classified as 'expected' in 54 cases (57%), 'paradoxical' in 17 cases (18%) and 'no effect' (Δ night ICP < 1 mmHg) in 23 cases (24%). There was a significant correlation between the Δ night ICP and the magnitude of valve setting change in both the investigated valves (Miethke ProGAV, p = 0.01 and Medtronic Strata, p = 0.02).

CONCLUSIONS

Paradoxical ICP changes can occur after shunt valve setting adjustments. This observation should be taken into account when performing ICP-guided valve adjustments and is highly relevant for the future development of "smart" shunt systems.

Для цитирования: D'Antona Linda, Craven Claudia Louise, Merchan Melida Andrea Jaime, Thompson Simon David, Bremner Fion, Thorne Lewis, Matharu Manjit Singh, Watkins Laurence Dale, Toma Ahmed Kassem. Paradoxical response of intracranial pressure to shunt valve setting adjustments.. Acta neurochirurgica. 2020. DOI: 10.1007/s00701-020-04462-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.