DOI 10.1016/s0266-7681(98)80176-9

Медицинская библиотека · Journal of hand surgery (Edinburgh, Scotland) · 1998

A biomechanical analysis of intrascaphoid compression using the Herbert scaphoid screw system. An in vitro cadaveric study.

Lo I K, King G J, Milne A D, Johnson J A, Chess D C

СтатьяПеревод в очереди

Аннотация

The purpose of this study was to evaluate the compressive capabilities of the Herbert scaphoid screw system. A transverse osteotomy was performed at the waist of eight scaphoids removing a 3 mm wafer of bone. A custom-designed load washer was inserted within this defect to measure intrascaphoid compression during Huene guide application, during screw insertion, after screw insertion and after guide release. The intrascaphoid compression was noted to peak during screw insertion; however it dramatically decreased without any clinical subjective evidence of decreased insertional torque by the surgeon. The average final compression as a percentage of maximal compression was 38 (SD 26) %.

Для цитирования: Lo I K, King G J, Milne A D, Johnson J A, Chess D C. A biomechanical analysis of intrascaphoid compression using the Herbert scaphoid screw system. An in vitro cadaveric study.. Journal of hand surgery (Edinburgh, Scotland). 1998. DOI: 10.1016/s0266-7681(98)80176-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.