Correcting common OCT artifacts enhances plaque classification and identification of higher-risk plaque features.

Jessney Benn, Chen Xu, Gu Sophie, Brown Adam, Obaid Daniel, Costopoulos Charis, Goddard Martin, Shah Nikunj, Garcia-Garcia Hector, Onuma Yoshinobu, Serruys Patrick, Hoole Stephen P, Mahmoudi Michael, Roberts Michael, Bennett Martin

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Optical coherence tomography (OCT) is used widely to guide stent placement, identify higher-risk plaques, and assess mechanisms of drug efficacy. However, a range of common artifacts can prevent accurate plaque classification and measurements, and limit usable frames in research studies. We determined whether pre-processing OCT images corrects artifacts and improves plaque classification.

METHODS

We examined both ex-vivo and clinical trial OCT pullbacks for artifacts that prevented accurate tissue identification and/or plaque measurements. We developed Fourier transform-based software that reconstructed images free of common OCT artifacts, and compared corrected and uncorrected images.

RESULTS

48 % of OCT frames contained image artifacts, with 62 % of artifacts over or within lesions, preventing accurate measurement in 12 % frames. Pre-processing corrected >70 % of all artifacts, including thrombus, macrophage shadows, inadequate flushing, and gas bubbles. True tissue reconstruction was achieved in 63 % frames that would otherwise prevent accurate clinical measurements. Artifact correction was non-destructive and retained anatomical lumen and plaque parameters. Correction improved accuracy of plaque classification compared against histology and retained accurate assessment of higher-risk features. Correction also changed plaque classification and prevented artifact-related measurement errors in a clinical study, and reduced unmeasurable frames to <5 % ex-vivo and ~1 % in-vivo.

CONCLUSIONS

Fourier transform-based pre-processing corrects a wide range of common OCT artifacts, improving identification of higher-risk features and plaque classification, and allowing more of the whole dataset to be used for clinical decision-making and in research. Pre-processing can augment OCT image analysis systems both for stent optimization and in natural history or drug studies.

Для цитирования: Jessney Benn, Chen Xu, Gu Sophie, Brown Adam, Obaid Daniel, Costopoulos Charis, Goddard Martin, Shah Nikunj, Garcia-Garcia Hector, Onuma Yoshinobu, Serruys Patrick, Hoole Stephen P, Mahmoudi Michael, Roberts Michael, Bennett Martin. Correcting common OCT artifacts enhances plaque classification and identification of higher-risk plaque features.. Cardiovascular revascularization medicine : including molecular interventions. 2024. DOI: 10.1016/j.carrev.2024.06.023.