A comparison study between the Siemens ADVIA 120 and manual method for the differential white blood cell count in goats.

Oikonomidis Ioannis L, Brozos Christos, Tsouloufi Theodora K, Kiossis Evangelos, Kritsepi-Konstantinou Maria

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Although widely used, the ADVIA 120 hematology analyzer has not been previously validated for determining the differential leukocyte count in goats.

OBJECTIVES

The aim of this study was to compare the differential leukocyte counts provided by the ADVIA 120 (A-diff) and the manual method (M-Diff) in goats.

METHODS

EDTA blood samples that were analyzed within 4 h of collection were used in the study. The following exclusion criteria were applied: inappropriately filled tubes or tubes containing clots, erroneous ADVIA peroxidase cytograms, and blood smears of poor quality. The A-Diff was compared with the M-Diff performed by two independent observers on 200 leukocytes.

RESULTS

Forty samples were included after previously excluding eight samples. The correlation between the A-Diff and M-Diff was very strong for eosinophils (r = .870, p < .001) and strong for lymphocytes (r = .796, p < .001) and neutrophils (r = .730, p < .001), while no significant correlation was observed for monocytes (r = .026, p = .872). The Passing-Bablok regression analyses revealed statistically significant constant errors for neutrophils (5.83%; 95% confidence interval [CI]: 0.41%, 12.18%) and eosinophils (1.89%; 95% CI: 1.17%, 2.71%). Bland-Altman analyses showed a statistically significant negative bias for lymphocytes (-5.0%) and a statistically significant positive bias for eosinophils (2.2%). The very low basophil percentages precluded a meaningful method comparison.

CONCLUSIONS

The ADVIA 120 overall demonstrated good performance for the differential WBC count in goats under the conditions of this study. Therefore, it can be considered suitable for routine hematologic screening in goats. Nonetheless, it should be emphasized that any abnormal result should be confirmed with a blood smear evaluation.

Для цитирования: Oikonomidis Ioannis L, Brozos Christos, Tsouloufi Theodora K, Kiossis Evangelos, Kritsepi-Konstantinou Maria. A comparison study between the Siemens ADVIA 120 and manual method for the differential white blood cell count in goats.. Veterinary clinical pathology. 2024. DOI: 10.1111/vcp.13337.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.