Validity of transit time-based blood pressure measurements in patients with and without heart failure or pulmonary arterial hypertension across different breathing maneuvers.

Becker Sara, Spiesshoefer Jens, Brix Tobias, Tuleta Izabela, Mohr Michael, Emdin Michele, Boentert Matthias, Giannoni Alberto

СтатьяПеревод в очереди

Аннотация

PURPOSE: Pulse transit time (PTT) derived by ECG and plethysmographic signal can be a promising alternative to invasive or oscillometry-based blood pressure (BP) monitoring in sleep laboratories because it does not cause arousals from sleep. Therefore, this study assessed the validity of PTT for BP monitoring under sleep laboratory-like conditions.

METHODS

Ten volunteers (55.8 ± 19.6 years), 12 patients with heart failure with reduced ejection fraction (HFrEF; 67.3 ± 8.6 years), and 14 patients with Nizza class I pulmonary arterial hypertension (PAH; 59.5 ± 13.4 years) performed different breathing patterns to simulate nocturnal sleep-disordered breathing (SDB). BP was measured at least every 15 min over 1 h using oscillometry (Task Force Monitor™) and PTT (SOMNOscreen™) devices in free breathing conditions and during SDB simulation (alternating phases of hyperventilation and apneas).

RESULTS

One hundred forty-two points of measurements were collected. No difference was found in both mean systolic BP (SBP) and diastolic BP (DBP) between oscillometric PTT-based BP measurements in the whole population and throughout the whole recording (SBP 111.3 ± 15.1 mmHg versus 110.0 ± 14.7 mmHg, p = 0.051; DBP 69.9 ± 12.2 versus 69.9 ± 14.2 mmHg, p = 0.701). Likewise, no significant difference in SBP and DBP was found between the two methods in the subgroups of healthy subjects, HFrEF patients and PAH patients, both in free breathing conditions (p > 0.05) and during SDB simulation (p > 0.05).

CONCLUSIONS

When monitoring BP in healthy subjects, and in patients with HFrEF or PAH, PTT provides a BP estimation comparable with oscillometric measurement, though slightly inaccurate, both in the condition of regular and unstable breathing.

Для цитирования: Becker Sara, Spiesshoefer Jens, Brix Tobias, Tuleta Izabela, Mohr Michael, Emdin Michele, Boentert Matthias, Giannoni Alberto. Validity of transit time-based blood pressure measurements in patients with and without heart failure or pulmonary arterial hypertension across different breathing maneuvers.. Sleep & breathing = Schlaf & Atmung. 2019. DOI: 10.1007/s11325-019-01848-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.